background image

Dansensor A/S

COPYRIGHT ©

PBI-320297-B 02/2014

CheckPoint O

2

3

CheckPoint O

2

 specifications

General:

• Min. 2000 measurements on new alkaline 

batteries.

• Sample flow 90 - 140 ml/min standard.
• Auto power-off after 1 minute.
• Memory function for 10 measurements (also 

after battery replacement).

• Display hold function (also after power off).
• Charging time for NiMH-battries: 3 hours 

(auto off)

• Operational temperature: 0 - 40°C
• Humidity: <95% RH (non-condensing)

O

2

 Sensor:

• Electrochemical
• Range: 0 - 100%
• Resolution: 0.1%
• Accuracy: 0.25% abs. + 2% rel.
• Temperature compensation:

< 2% rel. (steady state)

• Response time (T

95

):   15 sec.

• Expected sensor lifetime: 9 months

Recommended calibration interval: 6 months

Measuring system:

• Measuring time: 15 sec.
• Sample volume: approx. 23 ml
• Built-in pump max. -200 mbar vacuum, 

expected life-time > 500 hours (120.000 
measurements)

• Calibration: Air (20.9 % O

2

)

Cabinet:

• Weight: 0.5 kg
• Size: 74 x 63 x 135 mm (W x H x D)
• Class: IP53
• Material: Anodized aluminium

 DANSK................................................................................................5

 ENGLISH ...........................................................................................11

 ITALIANO..........................................................................................17

 DEUTSCH..........................................................................................23

 FRANÇAIS ........................................................................................29

 ESPAÑOL ..........................................................................................35

 PORTUGUÊS .....................................................................................41

 SVENSKA..........................................................................................47

 

РУССКИЙ

.....................................................................................53

DA

EN

IT

DE

FR

ES

PT

SV

RU

Содержание CheckPoint O2

Страница 1: ...PBI 320297 B 02 2014 CheckPoint O2 DE DA SV RU FR ES IT EN PT...

Страница 2: ...S COPYRIGHT PBI 320297 B 02 2014 CheckPoint O 2 Distributor Sales Service and Calibration Topac Inc 231 CJC Highway Cohasset MA 02025 Tel 781 7408778 Fax781 740 8779 Email sales topac com internet ww...

Страница 3: ...2 CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S 7 13 5 14 6 4 1 3 2 7 8 8 9 11 12 10...

Страница 4: ...condensing O2 Sensor Electrochemical Range 0 100 Resolution 0 1 Accuracy 0 25 abs 2 rel Temperature compensation 2 rel steady state Response time T95 15 sec Expected sensor lifetime 9 months Recommen...

Страница 5: ...10 Brugerguide DA CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTER...

Страница 6: ...second measuring period is started Reduced measuring time The measuring period can be reduced by pressing while measuring is in progress Memory function CheckPoint remembers up to 10 measurements incl...

Страница 7: ...n performing consecutive measurements in areas with large differences in the O2 concentrations the device will reach 95 of the true value during the first measurement 15 sec Example See illustration b...

Страница 8: ...During calibration the display alternately shows CAL and 20 9 for 45 seconds CheckPoint has now been calibrated Stop calibration Press to interrupt the calibration process External power supply charg...

Страница 9: ...t the needle retainer 9 in the opposite side and press it down until flush with the side plate 6 Test control Switch on the CheckPoint and start measuring with atmospheric air 20 9 Block the gas inlet...

Страница 10: ...205 Needle kit 0 5 mm 100 pcs Ser Cpl 280167 Needle kit blunt 10 pcs Ser Cpl 310335 Filter sample gas 0 2 water trap 10 pcs Ser Cpl Spare parts 200720 Pump CheckPoint Ser Cpl 200721 Kit hoses internal...

Страница 11: ...16 User Guide EN CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTES...

Отзывы: