DM DOC 040 Rev 1.0
LNCS
TM
ADHESIVE SENSORS
(LNCS sensors must be used in conjunction with Red LNC or Rainbow patient cables)
SENSOR
Weight
Range
Saturation
Accuracy
Pulse Rate
Accuracy
Low
Perfusion
Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNCS Adtx
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS Pdtx
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS Inf-L
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS Neo-L < 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
> 40 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS NeoPt-
L
< 1 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
SENSOR ACCURACY
Refer to Section 7,
Specifications
for SpO2, SpMet, SpCO and pulse rate accuracy. Unless
otherwise specified in the previous tables:
The sensor accuracy specified is when used with Masimo Rainbow SET technology Pulse
CO-Oximetry monitors or with licensed Masimo SET pulse oximetry modules during no
motion. Accuracy range variation equals ± 1 standard deviation. Plus or minus one
standard deviation represents 68% of the population. SpO2 accuracy represents from 70%
to 100%, pulse rate accuracy represents from 25 to 240 bpm, SpCO accuracy represents
from 1% to 40% and SpMet accuracy represents from 1% to 15%.
© 2013 DanMedical Ltd