2.8 Grounding Shielded Control Cables
Shield all control cables and connect the shield with cable
clamps at both ends to the metal cabinet.
shows
ground cabling for optimal results.
NOTE!
CT wires must be shielded or twisted pair to reduce noise
impact on measured signal.
Correct grounding control cables and cables for serial communication must be fitted with cable
clamps at both ends to ensure the best possible electrical connection.
Incorrect grounding Do not use twisted cable ends (pigtails) since these increase shield
impedance at high frequencies.
Ground potential protection When the ground potential between the filter and the PLC or other
interface device is different, electrical noise may occur that can disrupt the entire system. This
can be resolved by fitting an equalizing cable next to the control cable. Minimum cable cross
section is 8 AWG.
50/60 Hz ground loops When using very long control cables, 50/60 Hz ground loops may occur
that can disturb the entire system. This can be resolved by connecting one end of the shield
with a 100 nF capacitor and keeping the lead short.
Serial communication control cables Low frequency noise currents between filters can be
eliminated by connecting one end of the shielded cable to filter terminal 61. This terminal
connects to ground through an internal RC link. It is recommended to use twisted-pair cables to
reduce the differential mode interference between conductors.
Table 2.11 Grounding Shielded Control Cables
Operator Interface and Acti...
VLT Advanced Active Filter AAF006 D and E Frames Service Manual
MG90Z122 - VLT
®
is a registered Danfoss trademark
2-11
2
2