Danatronics, Corp.
18
To set the ball size on the MTG-99, follow these steps:
1.
Press the
Menu OK
key to access the Main Menu
2.
Scroll down to Measurements then press
Menu Ok
to select
3.
In the Measurements Menu scroll down to ball size and use the
right arrow
to
toggle to the right size of the screen where the ball size is listed
4.
Use the up and down arrows to highlight the desired ball size and press Menu Ok
to select.
5.
Press
F3
to go to the measurement screen or
F1
to go back to the main menu
Note:
The number in parenthesis is the maximum thickness for a given ball size. Shim
sizes may be slightly larger than the max value ie 0.241” vs 0.240”
3 Calibrating the Gage
The standard multipoint calibration should be completed every time the gage is being
used. It should be done after waiting 15 minutes for the gage and probe to warm up in
the environment in which they will be used. On thicker parts or if temperature changes
by more than 1-2 degrees F it is recommended to perform a Q-Cal. Q-Cal is used for
offset calibration during the measurement process to realign the calibration. We
recommend performing a Q-Cal every 15 minutes while the gage is in use.
Note:
It is strongly recommended to do a Q-CAL every 15 minutes, particularly on parts
thicker than .080” or 2mm. The MTG 99 will detect temperature changes which will lead
to errors and prompt the user to press Q-cal either on the unit above F4 or on the side
of the probe.