Shad-o-Box 1K/2K HS User Manual 

 

p. 12 

 

Optimizing Settings 

 
The performance of the gigabit Ethernet connection on the host computer generally can be improved by adjusting 
some settings in the NIC. The below advices refer to the Intel Gigabit CT desktop adaptor. For a more detailed 
explanation please refer to “Network Optimization Guide (PDF)” and “Network Imaging User’s Manual (PDF)” 
which is part of the Sapera LT software package. 
To reduce the interrupt overhead in the host the following settings in the NIC are advised. 
 

Packet Size 

For a good Gigabit Ethernet connection with minimal packet resend conditions, host computer performance can 
generally be improved by increasing the data packet size. Each streaming video packet causes an interrupt in the 
host computer. Therefore increasing the packet size reduces the CPU overhead required to handle video data from 
the GigE Vision detectors. 
A standard packet can have a size up to 1500 bytes. Many network cards support a jumbo packet mode that can 
extend that size up to 16Kbytes. In theory, a packet could be as large as 16 KB, but the CRC (cycle redundancy 
check) containing the checksum of each packet is not as efficient when the packet size grows larger than 9000 bytes. 
For this reason 

we advise to set the packet size to 9014 bytes

 (jumbo packet enabled). 

 

 

Interrupt Moderation 

Normally, each time a packet is received by the network card, the associated driver will receive an interrupt. 
Obviously, when the packet rate is very high (that is, at high transfer rate which is common for GigE Vision 
systems), this represents significant overhead. Most network cards have introduced an interrupt moderation mode 
where the card waits to have received a certain number of packets over a maximum period of time before issuing the 
interrupt. This helps reduce the burden on the CPU as it can process multiple packets during the same interruption. 
The Intel Gigabit CT Network adapter provides a configuration parameter to manually adjust the NIC interrupt 
moderation rate. By default the NIC driver sets this to Adaptive where the interrupt rate automatically balances 
packet transmission interrupts and host CPU performance. In most cases no manual optimization of the Interrupt 
Moderation Rate parameter is required.  
In some conditions, video frames from the GigE Vision detector may be transferred to the host display or memory 
buffer as data bursts instead of a smooth continuous stream. The NIC may be over-moderating acquisition interrupts 
to avoid over-loading the host CPU with interrupts. If priority is required for acquisition transfers (i.e. a more real-

Содержание Shad-o-Box 2K HS

Страница 1: ...Teledyne DALSA Shad o Box 1K 2k HS Detector User Manual December 11 2018 www TeledyneDALSA com...

Страница 2: ...it and electronics technology software and highly engineered semiconductor wafer processing Products and services include image sensor components electronic digital cameras X ray detectors vision proc...

Страница 3: ...ody and Front Cover 7 Periodic Maintenance 7 Calibration 7 5 Specifications 8 6 Connectors Cables and Status Indicators 9 7 Setting up the Detector 10 GigE Network Adapter 10 Connect Ethernet cable 10...

Страница 4: ...Shad o Box 1K 2K HS User Manual p 4 Shuttering 17 11 Mechanical Drawings 18 Revision History 20...

Страница 5: ...ALSA assumes no responsibility for the proper installation of X ray shielding X ray shield enclosure testing or safe and prudent operation of the camera system in the end user s installation It is the...

Страница 6: ...ld by Teledyne DALSA as replacement parts may result in increased emission or decreased immunity of the Shad o Box HS Detector Static precautions Observe proper ESD static control procedures when hand...

Страница 7: ...tector body surface Wipe off with a clean soft cloth Do not use any harsh cleansers or solvents that may damage the paint or other finishes on the detector In the event of detector exposure to bio haz...

Страница 8: ...Non linearity 10 90 FS 1 5 Readout period 3 45 ms Output data rate 40 MHz 1 dark current doubles approx every 8 C 2 ADU Analog Digital Unit 1 LSB Least Significant Bit 3 time required to transfer ima...

Страница 9: ...ype FHG 0B 302 CLAD52Z straight type FGG 0B 302 CLAD52 Detector Status LED The detector is equipped with a color LED used to display the status of the detector s operation The table below summarizes t...

Страница 10: ...Intel PRO 1000 MT adapter Review the NIC documentation concerning any special driver required for Windows Install the PCI bus Gigabit NIC as described by the NIC manufacture s documentation Connect Et...

Страница 11: ...on Sapera LT and CamExpert Installation A link to download the software will be emailed to you after submitting the application Complete the download process and run the installation file e g SaperaLT...

Страница 12: ...rger than 9000 bytes For this reason we advise to set the packet size to 9014 bytes jumbo packet enabled Interrupt Moderation Normally each time a packet is received by the network card the associated...

Страница 13: ...conditions the host PC system CPU may be busy with tasks other than the imaging application Incoming image packets remain in the PC memory allocated to store packets instead of immediately being copi...

Страница 14: ...used to manage flow control i e the speed at which stream packet can be output to the network This is useful when connecting multiple detectors to the same port of the network card or when the networ...

Страница 15: ...equests will not be used Connecting 1 After the install is complete plug in the detector communication cable 2 Power up the detector 3 After powering up the detector the application automatically dete...

Страница 16: ...he grabbing process generates multiple frames by operating in a continuous mode To snap an image Click the Snap button The Detector has now snapped a single frame The image appears in the GUI window T...

Страница 17: ...interval is marked with the dark green rectangle Row Pixel Array Time ReadOut Integration Time Frame N Frame N 1 Trigger Input Trigger Output The period of time in between successive frame read outs...

Страница 18: ...Shad o Box 1K 2K HS User Manual p 18 11 Mechanical Drawings Shad o Box 1K HS Model all measurements are in mm...

Страница 19: ...Shad o Box 1K 2K HS User Manual p 19 Shad o Box 2K HS Model all measurements are in mm...

Страница 20: ...ox 1K 2K HS User Manual p 20 Appendix A Revision History Revision Number Change Description Revision Date 01 Draft release May 10 2016 02 New address January 12 2018 03 Added SB2kHS detector December...

Отзывы: