Figure 34 A histogram of a black image.
Step 3: Select
Calibration
from Flat Field Correction Mode drop-down menu.
Figure 35: FFC Calibration Selection
Step 4: Select
2048
or
4096
option from Flat Field Calibration Sample Size. The 2048 option
reduces calculating time; 4096 option returns result that is more accurate.
Step 5: Press the Press… command in the Calibrate FPN menu