Dakota Ultrasonics
18
Scan Mode
While the
DX-1
excels at making single point measurements, it is
sometimes desirable to examine a larger region, searching for the thinnest
point. The
DX-1
includes a feature, called Scan Mode, which allows it to do
just that.
In normal operation, the
DX-1
performs and displays four
measurements every second, which is quite adequate for single
measurements. In Scan Mode, however, the tool performs sixteen
measurements every second, but does not display them. While the
transducer is in contact with the material being measured, the
DX-1
is
keeping track of the lowest measurement it finds. The transducer may be
"scrubbed" across a surface, and any brief interruptions in the signal will be
ignored. When the transducer loses contact with the surface for more than
a second, the
DX-1
will display the smallest measurement it found.
When the
DX-1
is not in calibration mode, press the
UP
arrow key to
turn Scan Mode on and off. A brief message will appear in the display
confirming the operation. While scanning, the display will show a moving
series of dashes instead of a thickness value. When the transducer is
removed from the material being scanned, the
DX-1
will (after a brief
pause) display the smallest measurement it found.
Содержание DX-1
Страница 4: ...Dakota Instruments blank page ...
Страница 6: ...Dakota Ultrasonics 2 blank page ...
Страница 18: ...Dakota Ultrasonics 14 ...
Страница 27: ...DX 1 Ultrasonic Micrometer 23 blank page ...
Страница 31: ...DX 1 Ultrasonic Micrometer 27 blank page ...
Страница 33: ...DX 1 Ultrasonic Micrometer 29 blank page ...
Страница 35: ...DX 1 Ultrasonic Micrometer 31 blank page ...