Dakota Ultrasonics
12
Probe Zero
Setting the Zero Point of the
DX-1
is important for the same reason that
setting the zero on a mechanical micrometer is important. If the tool is not
"zeroed" correctly, all of the measurements the tool makes will be in error
by some fixed number. When the
DX-1
is "zeroed", this fixed error value is
measured and automatically corrected for in all subsequent measurements.
The
DX-1
may be "zeroed" by performing the following procedure:
Performing a Probe-Zero
1) Make sure the
DX-1
is on.
2) Plug the transducer into the
DX-1
. Make sure that the connectors
are fully engaged. Check that the wearface of the transducer is
clean and free of any debris.
3) On the face of the
DX-1
, below the keypad, is the metal probe-disc.
Apply a single droplet of ultrasonic couplant to the face of this disc.
4) Press the transducer against the probe-disc, making sure that the
transducer sits flat against the surface of the probe-disc. The
display should show some thickness value, and the Stability
Indicator should have nearly all its bars illuminated.
5) While the transducer is firmly coupled to the probe-disc, press the
PRB-0
key on the keypad. The
DX-1
will display "Prb0" while it is
calculating its zero point.
6) Remove the transducer from the probe-disc.
Содержание DX-1
Страница 4: ...Dakota Instruments blank page ...
Страница 6: ...Dakota Ultrasonics 2 blank page ...
Страница 18: ...Dakota Ultrasonics 14 ...
Страница 27: ...DX 1 Ultrasonic Micrometer 23 blank page ...
Страница 31: ...DX 1 Ultrasonic Micrometer 27 blank page ...
Страница 33: ...DX 1 Ultrasonic Micrometer 29 blank page ...
Страница 35: ...DX 1 Ultrasonic Micrometer 31 blank page ...