__________________________________________________________________________________
23
HOW TO COMPENSATE LIQUID SAMPLE TEMPERATURE
PREPARATION 4 - If the temperature compensation is necessary, follow the procedure below to change the
compensated liquid temperature and the compensating temperature rate of liquid sample.
What is "the compensating temperature rate"- The changing number of the density when the liquid is raised by1
degree centigrade. The user should put the coefficient manually.
Importance:
When the liquid used for measurement is at a sub-normal temperature then this feature enables the
user to enter this Temperature for the liquid so that the density measurement is calculated compensating this
difference in temperature.
Configuration Menu
Temperature Compensate
Press “Tare + Key A” to enter the “Configuration Menu”
Go to “Temperature Compensate” by using keys “A” or “B” till you reach this Option.
There are two options here Temperature compensate for:
-Water
-Any other liquid as per the user
For Water:
After entering the Temperature Compensate mode, click “Enter” on Water
Enter the values in “Water Actual Temperature” by using keys “A” or “B” and press “ENTER”
Enter the values in “Water Compensate Temperature” by using keys “A” or “B” and press “ENTER”
For any other Liquid:
After entering the Temperature Compensate mode, click “Enter” on Liquid Temperature Compensation
Enter the values in “Liquid Actual Temperature” by using keys “A” or “B” and press “ENTER”
Enter the values in “Liquid Compensate Temperature” by using keys “A” or “B” and press “ENTER”