the PATIENT or other leads is avoided. Temporary unused ACTIVE ELECTRODES should be
stored so that they are isolated from the PATIENT.
10) For surgical procedures where the h.f. current flow through parts of the body having a relatively
small cross-sectional area, the use of bi-polar techniques may be desirable in order to avoid
unwanted coagulation.
11) Apparent low output or failure of the HF SURGICAL EQUIPMENT to function correctly at the
normal operating settings may indicate faulty application of the NEUTRAL ELECTRODE or poor
contact in its connections. In this case the application of the NEUTRAL ELECTRODE and its
connections should be checked before selecting higher power.
12) The supplier will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the USER's appropriately qualified
technical personnel to repair those parts of EQUIPMENT which are designed by the
manufacturer as repairable.
13) It may cause a malfunction when the unit is used in an environment that has strong magnetic
fields i.e. MRI room.
14) Do not use the unit other than your intended purpose. It may cause unexpected side effects.
15)
Attention is called to the danger of ignition of endogenous gases
Sales &
Service Center
Address:
Tel :
Fax :
Website:
3, Yeoksam-ro 33-gil, Gangnam-gu, Seoul, Korea
+82-2-558-1711
∼
6
+82-2-554-0317
www.daiwha.com
Address:
Tel:
mdi Europa GmbHLangenhagener Str. 71
D-30855 Hannover-Langenhagen, Germany
+49-511-3908 9530
Содержание Meditom DT-300S
Страница 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches...
Страница 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage...
Страница 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S...
Страница 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage...
Страница 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S...
Страница 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage...
Страница 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S...
Страница 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage...
Страница 47: ...9 Load Regulation Vpeak Graphs47 9 5 Forced Soft Coagulation Mode 9 5 1Load Regulation DT 400S DT 300S...
Страница 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage...