6
. Operating and Sequence33
6) When using T.B.H, connect with the Monopolar Outlet1.
7) When using the foot switch, connect with the Mono-polar Foot Switch inlet located at the rear
panel and use the Monopolar Outlet1 or Monopolar Outlet2.
8) After finishing surgery, Turn OFF the Power Switch.
TIP What is Output Path Selector?
1) This slide switch alters the path of RF Output by Mono-polar Foot Switch. RF Output will go out
to Monopolar Outlet1 When the slide switch is selected Mono Outlet 1.
2) This slide switch alter the path of RF Output by Mono-polar Foot Switch. RF Output will go out
to Monopolar Outlet 2 When the slide switch is selected Mono Outlet 2.
Содержание Meditom DT-300S
Страница 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches...
Страница 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage...
Страница 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S...
Страница 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage...
Страница 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S...
Страница 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage...
Страница 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S...
Страница 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage...
Страница 47: ...9 Load Regulation Vpeak Graphs47 9 5 Forced Soft Coagulation Mode 9 5 1Load Regulation DT 400S DT 300S...
Страница 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage...