background image

Q-208-081 

 

 
 
2.5 CE marking (Only auxiliary supply DC24V specification product.) 

(1) Conformity technical standard 

・EMC conformity technical standard 

EMI(Emission)  EN61000-6-4 
EMS(Immunity)  EN61000-6-2 

・Safety technical standard 

EN61010-1  CATⅢ(Maximum circuit voltage, 300V),Pollution degree 2 

 

(2) Specification according to EMC test item 

EMC test item 

Limits 

Conformity 

technical standard 

EMI 

Electromagnetic  
radiation 
disturbance 

Frequency band 

Quasi-peak 10m distance 

EN61000-6-4:2001 
EN55011:1998 
classA,Group1 

30 ~ 230MHz 

40dB(μV/m) 

230 ~ 1000MHz 

47dB(μV/m) 

EMS 

Electrostatic 
discharge 

Contact discharge 
  ±4kV (Charge voltage) 
Air discharge 
  ±8kV (Charge voltage) 

After test, 
  

Output variation

 ±0.5% 

EN61000-6-2:2005 
EN61000-4-2:1995 

Radio-frequency 
amplitude modulated 
electromagnetic 
field 

Frequency 

Field strength 

During test: 
  

Output variation

 ±20% 

After test: 
  

Output variation

 ±0.5% 

EN61000-6-2:2005 
EN61000-4-3:2006 

80 ~ 1000MHz  10V/m,80% AM(1kHz) 

1.4 ~ 2.0GHz  3V/m,80% AM(1kHz) 

2.0 ~ 2.7GHz  1V/m,80% AM(1kHz) 

Fast transients 

Power port :Line to earth ±2.0kV 
Signal port:Line to earth ±1.0kV 

After test: 
  

Output variation

 ±0.5% 

EN61000-6-2:2005 
EN61000-4-4:2004 

Surge 

Power port :Line to line  ±0.5kV 
Signal port:Line to earth ±0.5kV 
Signal port:Line to earth ±1.0kV 

After test: 
  

Output variation

 ±0.5% 

EN61000-6-2:2005 
EN61000-4-5:2006 

Radio-frequency 
common mode 

Frequency:0.15 ~ 80MHz 
Voltage level:10V 
80% AM(1kHz) 

During test: 
  

Output variation

 ±20% 

After test: 
  

Output variation

 ±0.5% 

EN61000-6-2:2005 
EN61000-4-6:1996 

Power-frequency 
magnetic field 

Frequency:50/60Hz 
Field strength:30A/m 

During test: 
  

Output variation

 ±20% 

After test: 
  

Output variation 

±0.5% 

EN61000-6-2:2005 
EN61000-4-8:1994 

 

(3) Installation environment 

Over voltage Category Ⅲ 

 
 

Содержание HSQT2-93A

Страница 1: ...Q 208 081 INSTRUCTION MANUAL HIGH SPEED MULTI POWER LINE TRANSDUCER HSQT2 93A...

Страница 2: ...ammables combustible chemicals gas to prevent fire Connect specified power supply Connecting power supply beyond specification causes fire product failure When dust is on the terminal wipe it off afte...

Страница 3: ...8 3 Handling explanation 3 1 Outline dimension 9 3 2 Mounting instruction 9 3 3 The example of combination mounting dimension 9 3 4 Installation and removal to a rail 10 3 5 Connection 11 3 6 Handlin...

Страница 4: ...measurement range selection change The response time is high speed Current Voltage Active power Reactive power Less than input 1 cycle 10ms 99 Frequency Less than 1 second 99 As an option hydrogen su...

Страница 5: ...4 DC0 10V Over 2k 5 DC1 5V Over 600 8 DC 5V Over 600 9 DC 10V Over 2k B DC4 20mA Below 550 Auxiliary supply specification Mark Auxiliary supply 1 AC85 253V 50 60Hz Rated voltage 100 110V 200 220V DC8...

Страница 6: ...5Hz 0 5 Zero cross cycle arithmetic Voltage V RS or V RN is measured It can alter into the measuring range of 1 3 with front DIP switch About measurement 1 A class index is a rated frequency 50 60Hz 1...

Страница 7: ...ss index Update time of measured value 5ms Response time Time within 1 of final constant value when step input is charge A V W var With no moving average Less than input one cycle 10ms 2 times of movi...

Страница 8: ...and it applies for 5 minutes within 10 of errors Auxiliary supply circuit Voltage circuit Normal Common Over 1 5kV Current circuit Common Over 1 5kV Analog output Induction Over 1 0kV 3 Electromagnet...

Страница 9: ...ated electromagnetic field Frequency Field strength During test Output variation 20 After test Output variation 0 5 EN61000 6 2 2005 EN61000 4 3 2006 80 1000MHz 10V m 80 AM 1kHz 1 4 2 0GHz 3V m 80 AM...

Страница 10: ...ew mounting In screw mounting please install with M4 screw or M5 screw However the screw is not attached The tightening torque of a screw M4 1 0 1 3 N m M5 2 0 2 5N m Please detach an interval lining...

Страница 11: ...n in a BOX bottom is inserted in a rail Next if it pushes against the lower side arrow it is fixed with a lower slider Removal from a rail A minus screwdriver is inserted in the square socket of a sli...

Страница 12: ...is product should perform the right usage with careful attention to the following item 1 Please check that the voltage of power supply and the input signal to supply match the specification of this pr...

Страница 13: ...utput V V 5 2 5 10 5 0 0 1 0 1 kW LEAD 1 0 LAG 1 kvar AC input DC output V mA 3 12 5 20 1 4 0 0 5 1 kW 1 0 1 kW LEAD 1 0 LAG 1 kvar AC input DC output 3 Frequency Input Output Input Output Input Outpu...

Страница 14: ...frequency can be altered with a front DIP switch DIP switch ON 1 OFF 0 S12 S11 S10 S9 S8 S7 S6 S5 S4 S3 S2 S1 W measurement range var measurement range S1 S2 S3 Input 110V 5A is 1A rating S4 S5 Input...

Страница 15: ...continuing pushing a switch a coarse adjustment becomes possible DOWN setting for output adjustable DOWN This switch drops the output for the selected target for adjustment By pushing a short time swi...

Страница 16: ...13 Hitotsuya 1 chome Adachi ku Tokyo 121 8639 Japan Phone 03 3885 2411 Fax 03 3858 3966 Kyoto Office 1 19 Ichinobe Nishikawahara Jyoyou shi Kyoto 610 0114 Japan Phone 0774 55 1391 Fax 0774 54 1353 Re...

Отзывы: