DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Forced Convection type Smart Oven 35/80
<Fig.3.3.31> setting offset values
-
Limit of offset values are -20.0
℃
~ +20.0
℃
, and current temperature after setting offset
value displays the value that includes the offset.
-
Once you set the offset value, the value is stored and the offset value is reflected in the
current temperature even you press ON/OFF button.
3.3.5. Error Message
-
If happen a error, Error notice will be showed in the display and if you click it, you can see
what happened in detail.
* When malfunctions in the Touch Screen occur, please press the Reset Button by
inserting an appropriate tool (such as a paper clip) into the Reset hole, which is located
in the right frame of the Touch Screen as shown in the figure below. The Touch Screen
is rerun by pressing the Reset Button.
ERROR
Message
Causation
Solution
Temp Error
When display show unreasonable
temp. value by out of order of temp.
sensor
Call the A/S Center.
Door Open
When over 10 Sec. after open the
Door.
Check status of the Door and if still
happen it even door closed, call the A/S
center.
Revolution
Error
When the Communication error
problem of equipment and Touch
Board.
Please check the connection of the Touch
Board.