      2.3 Product Specifications 

 

Capacity & Models 

50Lit 

ThermoStable

 TM

 

SIF-50 

105Lit 

ThermoStable

 TM

 

SIF-105 

155Lit 

ThermoStable

 TM

 

SIF-155 

Dimension 

(w x d x h) 

Interior(mm) 

370X350X419 

485X410X533 

550X475X600 

Exterior(mm) 

518X637X700 

635X695X841 

700X760X906 

Heater Power 

200W 

300W 

400W 

 

 

Temp. 

Range 

Ambient Tempe5℃  to 70℃ 

Fluctuation 

±0.2℃  at 37℃,  ±0.3℃  50℃ 

Variation 

±0.5℃  at 37℃,   

±0.9℃  at 50℃ 

±0.4℃  at 37℃,   

±0.7℃  at 50℃ 

Sensor 

PT 100 

Control Resolution 

±0.1℃ 

Heat-up Time 

17min. to 37℃,   

28min. to 50℃ 

17min. to 37℃,   

30min. to 50℃ 

Recovery Time 

(Door open 30sec) 

3min. to 37℃ 

4min. to 50℃ 

Controller 

Smart-Lab

TM

 Controller 

RS232 Port 

Available to connect with PC 

Display 

4"Full Touch Screen TFT LCD 

Timer 

99hr 59min (delay/continuous function) 

 

Material 

Internal 

Stainless Steel(#304) 

External 

Powder Coated Steel 

Inner Door 

Inner Tempered Glass Door 

Shelves 

2ea of PE-Coated Steel Wire Shelves included 

Ventilation 

1ea of Ventilation hole with Stainless Steel Cap, Hole Diameter 40mm 

Circulation 

Forced Air Type 

Safety device 

Over Temp. and Over Current Protector, Sensor Error Detector 

 

Others 

UL Certified, CE Certified 

Storage Function (Temp. and Time) 

Locking Mode (Jog-Dial Input Disabled) 

Alarm (Error Status and Timer-end) 

Net Weight (kg) 

46 

66 

81 

Packing Size (wxdxh)(mm) 

628X752X775 

710X731X1,018 

731X810X1,018 

Shipping Weight(kg) 

58 

74 

92 

Power Consumption 

216.3 W 

302.9 W 

438.1 W 

Energy Consumption, at 37℃ 

                                      , at 50℃ 

28 Wh 

43 Wh 

58 Wh 

84 Wh 

129 Wh 

145 Wh 

Power Supply 

120V, 60Hz or 230V, 50/60Hz 

 

 

Содержание ThermoStable SIF-50

Страница 1: ...Certified CE Certified PL Insurance Product Liability Insurance Validation IQ OQ Service User Manual Version 2 2 2 written for ThermoStable TM SIF 50 105 155 models ThermoStableTM SIF 105 Forced Conve...

Страница 2: ...DAIHAN Sci com or ALL forLAB com Shopping Mall DAIHAN Scientific Forced Convection type Smart Incubator 2 81...

Страница 3: ...n 5 1 3 Product Configuration 6 2 Installation 2 1 Product Installation 7 2 2 Installation of WiReTM App Web Service 9 2 3 Product Specifications 16 3 Product Usage 3 1 Precautions for Use 17 3 2 Name...

Страница 4: ...ith the meaning Warning Failure to follow the instructions may result in injury Caution Failure to follow the instructions may result in product failure Proximity Caution Keep clear of the product dur...

Страница 5: ...menus have been simplified through 4 inches Full Touch Screen by applying Smart LabTM Controller 4 It sends the unusual condition in real time by the self testing system of the experimental equipment...

Страница 6: ...onent Name No of Configuration ThermoStable TM SIF 50 105 155 1 unit PE Coated steel Wire Shelves 2 ea Including a copy of the Operation Manual Models ThermoStable TM SIF 50 Forced Convection type Sma...

Страница 7: ...or normal operation at a rated voltage so please check the voltage status before installing the product 7 When installing and using the product please certainly use the power cord provided with the pr...

Страница 8: ...symbols stickers Attached to side of product For meaning of each symbol please refer to Precautions Before Use Safety Instructions Product Liability PL Insurance Admission DAIHAN Scientific Co Ltd is...

Страница 9: ...t the factory please connect by finding the wireless LAN among the components of the product This description is for the installation method for users who want to use WiReTM App Web Service Only matte...

Страница 10: ...stem menu to go on the system setting Push the return button placed on top right of all displays for each menu to go back to previous display System Settings 1 Current Time Set The present time is set...

Страница 11: ...Fi Setup Set The WiReTM Service will only be available by the wireless internet connection Users should make the product be connected to internet via wireless network Click the Check Box of Wi Fi wind...

Страница 12: ...iRe Mode will not be switched normally Be sure to secure stable Wi Fi connection to the product After completing the internet connection users can find the Serial Number Model and Organization Name as...

Страница 13: ...esponding to users to be deleted therefrom 3 Registration of New User Users who are using the Smart Lab device and WiRe Service for the first time shall register users information Input the ID Passwor...

Страница 14: ...the list then they cannot be modified thereafter 4 Setting of Authority of each User Users registered in the list can be set as an administrator or as a general user The Admin the administrator is the...

Страница 15: ...of the message showing the disability to invoke the information of registered users may come from the unstable Wi Fi connection Check the status of Wi Fi connection and then reenter into the menu of W...

Страница 16: ...er Smart LabTM Controller RS232 Port Available to connect with PC Display 4 Full Touch Screen TFT LCD Timer 99hr 59min delay continuous function Material Internal Stainless Steel 304 External Powder C...

Страница 17: ...ge or distress the power cord Please do not hurt on the power cord or process it or force bend it or pull it and tie it Also if heavy objects are placed over the power cord or the power cord is insert...

Страница 18: ...sly shines and can get rained on it is prohibited It may cause overheating or electrical fault of the product Please do not use flammable or volatile products in the vicinity of the product to minimiz...

Страница 19: ...erlogged please contact the deliverer or our technical service engineer Do not operate the unit before checking with an engineer otherwise it may cause an electrical shock or a fire The Sound Level of...

Страница 20: ...ly Air Duct Ventilation of Chamber Controller Set and Check operation Window Safety Double Door Handle Door open and close Fig 3 2 1 Forced Convection Smart Incubator ThermoStableTM SIF 105 Fig 3 2 2...

Страница 21: ...s Description Power Switch Power ON OFF Fuse Protect unit from High Current Safety Switch Prevent overheat Fig 3 2 4 Side of ThermoStableTM SIF ThermoStable SIF 50 120V 230V 250V F 5AH ThermoStable SI...

Страница 22: ...the calibration temperature CHART Monitor the operating status by chart SETTINGS Set up and adjust various functions of the instrument Error If the error of the instrument occurs perform the alarm fu...

Страница 23: ...Setup window will be popped up wherein users may set desired temperature with the value inputted directly through the numeric Button Temperature setting is adjustable in units of 0 1 and can be set to...

Страница 24: ...ings 1 Running Mode After Power Failure It sets a mode for power is abruptly cut off and turned back up due to power supply issues such as power outage When setting a Stop mode It stands ready at the...

Страница 25: ...tomatically operates at a temperature set that it previously operated on and the operation time on main screen continues its counting B Program Mode It restarts the count from the remaining time of fi...

Страница 26: ...activated Set Fixed Operation Time Mode to ON and press Fixed Operation Time to set the value of timer It operates for a time it set when you start the product from main screen If Fixed Operation Tim...

Страница 27: ...perature after setting the segment time to reach the target temperature If the maintaining time for target temperature is not needed do not make second segment for the target temperature B Normal It o...

Страница 28: ...1 Key Lock The password enabled to operate the product can be locked to prevent unauthorized operation of the controller for particular settings of the product Slide the button to the right to activa...

Страница 29: ...ern can store up to 120 and a maximum of 100 segments can be stored in 1 Pattern Pattern is created by using button and in case you want to delete press the number of Pattern you want to delete and ve...

Страница 30: ...graph Enter the value of Segment after creating Pattern and Segment You can verify the temperature and time set for the Segment selected when you select Segment from the Segment window and touch You...

Страница 31: ...reated in the operation of Program Mode 1 Program button it enters into the Program setting screen rapidly 2 Step button It skips to next Segment while in operation 3 Hold button It temporarily stops...

Страница 32: ...times and can be set to infinity by pressing button at 1 2 You can edit the order of operation after setting the number of repeat Select Start Segment and End Segment by pressing button from Sequence...

Страница 33: ...rn can select from stored Pattern other than selected Pattern with Repeat operation it does not allow to operation in continuation of other Pattern after the Repeat operation if set to None Wait In op...

Страница 34: ...at internal memories at every 10 seconds and maximum 6 months of data are stored Please use Data Backup once every 6 months Temperature scope shown on graph can be changed using the set button on the...

Страница 35: ...e pressing the point detailed information of the location emerges on a detail display You can check the information on graph by moving at a small increment of the point using arrows on the detail disp...

Страница 36: ...art of Touch Screen Controller and if still happen the error please call the A S center When malfunctions in the Touch Screen occur please press the Reset Button by inserting an appropriate tool such...

Страница 37: ...in the vicinity of room temperature when slowly turning clockwise up to 120 after adjusting the Safety Knob to 0 3 If the power of the product is ON check the Tak sound in the vicinity of current temp...

Страница 38: ...at 5 C 40 C 3 If not being used for a long period unplug and dry before packaging for storage 4 If replacement parts are required due to the failures while in use it must use the original parts suppl...

Страница 39: ...lated Heating but temperature is fluctuated Check the temperature that is currently set Make sure that the power switch is in the ON position Check that SAFETY Knob is not placed in the 0 position and...

Страница 40: ...ackets for 50 Lit ONS050 w340xd310mm DH WON11105 Wire Shelf Stainless Steel with 2 Brackets for 105 Lit ONS105 w455xd375mm DH WON11155 Wire Shelf Stainless Steel with 2 Brackets for 155 Lit ONS155 w52...

Страница 41: ...all functions before shipment 2 This instrument is warranted for 2 years from purchasing date 3 A breakdown caused by customer s improper use or natural disaster is excepted from this warranty 4 This...

Отзывы: