8.Features
© 2023 China Daheng Group, Inc. Beijing Image Vision Technology Branch 156
Load parameters (UserSetLoad): Load the vendor default configuration parameters or the user
configuration parameters into the effective configuration parameters. After this operation is performed, the
effective configuration parameters will be covered by the loaded parameters which are selected by the
user, and the new effective configuration parameters are generated. The operation steps are as follows:
1)
Use UserSetSelector to select Default or UserSet0.
2)
Execute UserSetLoad command to load the User Set specified by UserSetSelector to the device and
makes it active.
Change startup parameter set (UserSetDefault): The user can use UserSetDefault to select Default or
UserSet0 as the UserSetDefault. When the camera is reset or powered on again, the parameters in the
UserSetDefault will be loaded into the effective configuration parameters.
8.3.9.
Device User ID
The MER-G-P/MER-G series camera provides programmable device user ID function, the user can set a
unique identification for the camera, and can open and control the camera by the unique identification.
The user-defined name is a string which maximum length is 16 bytes, the user can set it by the following
ways:
1)
Set by the IP Configurator, for details please see section 9.1:
Figure 8-26 IP Configurator
2)
Set by calling the software interface, for details please see the Programmer's Guide.
When using multi-cameras at the same time, it is necessary to ensure the uniqueness of the
user-defined name of each camera, otherwise, an exception will occur when the camera is
opened.