DAGE-MTI Investigater Скачать руководство пользователя страница 6

Operation 

 

Figure 4:  Dage-MTI 

InvestiGater 

Front Panel 

 

The 

InvestiGater's

 front panel allows the user to switch 

between the stored image and the incoming 

LIVE

 image.  The 

gating of the CCD camera is controlled through the front 
panel's 

GATE TIME

 switch and the 

SIG LEV

 indicator tells 

when the output video is the correct level. 
 

POWER 

The 

InvestiGater's

 power is turned on and off by this 

switch.  The green LED above the switch lights when the unit 
is on.  The 

InvestiGater

 provides a shunt video path through 

the unit even when the power is off.  When the 

InvestiGater 

is attached to the 

CCD100

, this switch controls power to both. 

 

GATE TIME 

This switch controls the time that the camera will be 

gated.  The unit allows gating times from 4 frames (1) to 
2048 frames (MAX).  The switch doubles the gating time for 
each position.  The gating cycle is initiated by the 

ACT

 front 

panel switch. 

 

Содержание Investigater

Страница 1: ...User s Manual DAGE MTI Investigater CCD Camera Gater Frame Grabber...

Страница 2: ...Purchaser s Record Model Name DAGE MTI Investigater Serial Number Dealer s Name Dealer s Address Dealer s Phone Number Date Purchased P O Number...

Страница 3: ...eras such as the CCD 300RC CCD 72AS and the IR 1000 models The InvestiGater and CCD Camera package provides a complete imaging system for any situation A superior image can be obtained regardless of t...

Страница 4: ...CCD 300RC The CCD 300RC needs an Aux Gating Cable Dage P N 835320 08 which is included with the cooled models Note in this configuration both units need a power supply connection Figure 3 shows the In...

Страница 5: ...ING P S CAMERA VID OUT GATE VID IN CCD300 HEAD POWER CABLE P N 835320 08 Figure 2 InvestiGater CCD 300RC Connection MONITOR InvestiGater CAMERA VID OUT POWER TABLE TOP SWITCHING P S 72AS POWER SUPPLY...

Страница 6: ...InvestiGater s power is turned on and off by this switch The green LED above the switch lights when the unit is on The InvestiGater provides a shunt video path through the unit even when the power is...

Страница 7: ...en the gate cycle has been completed the light turns off When in the RUN mode another gating cycle automatically begins The ACT button works in conjunction with the LIVE switch If the InvestiGater is...

Страница 8: ...s is in a series of snapshots that are updated each gating period This gating process can be halted by either pressing the ACT button or switching to the LIVE mode POWER IN This 5 pin DIN connector re...

Страница 9: ...ored to provide a constant current to the unit The center pin is positive polarity CAMERA This 4 pin MINI DIN connector provides the complete interconnection between the InvestiGater and the CCD100 Po...

Страница 10: ...w to high transition are saved in the InvestiGater s frame memory The InvestiGater s gate signal inhibits the readout of the CCD s image and allows light to gather in the CCD for the period of time de...

Страница 11: ...eo signal should comply with either EIA RS 170A 525 60 or CCIR 625 50 video signal standards VIDEO OUT The video output of the InvestiGater is available at this BNC connector The video has a source im...

Страница 12: ...ames in 2X intervals Video Input EIA RS 170A 525 60 1Vpp into 75 or CCIR 625 50 1Vpp into 75 Signal Level LED Yellow Light level or Gate Time too low Green Light level or Gate Time correct Red Light l...

Страница 13: ...o Picture Check or try the following 1 InvestiGater in LIVE 2 Camera Power On 3 InvestiGater CCD100 Cable Attached 4 Light Level Too Low 5 Camera Lens Open 6 Monitor Connected to VID OUT 7 Monitor Pow...

Страница 14: ...been subject to abuse neglect accident improper installation or on which the serial number has been removed or damaged Units that have been altered without the prior permission of DAGE MTI are not cov...

Страница 15: ...ems caused by or inflicted upon associated equipment such as digitizing systems video tape recorders cameras microscopes etc b Damage caused by accident misuse improper power source fire flood lightni...

Страница 16: ...DAGE MTI Inc 701 N Roeske Ave Michigan City IN 46360 219 872 5514 Fax 219 872 5559 E mail dage dagemti com http www dagemti com 970292 01 8 13 02...

Отзывы: