6
Introduction
Product Name
PURISMA
* SPOT handpiece is used for acne, scar treatment and lifting by concentrating plasma energy on
needle tip, while JET handpiece is used for toning, skin sterilization and regeneration by emitting high
plasma energy over a wide range of the skin.
Technical Specification
.
General
Power Input
……………………………………………………………………………………AC100-240V, 50/60Hz
Maximum Power Consumption
……………………………………………………………………………………50VA
Main Fuses
……………………………………………………………………………………250V, 3A (5 x 20 mm)
Size (height x width x depth)
……………………………………………………………………110 x 280 x 300 mm
Weight…………………………………………………………………………………………………………………3 kg
Output Parameter
Output Frequency……………………………………………………………………… Spot: 70KHz / JET: 12KHz
Output Modes………………………………………………………………………………………… SPOT / JET
Output Voltage
Program
Level
Output voltage
Remarks
SPOT
1 Level
3.00Vrms ±30%
Non-inductive resistance
500Ω
2 Level
3.10Vrms ±30%
3 Level
3.20Vrms ±30%
JET
1 Level
3.8Vrms ±30%
2 Level
3.8Vrms ±30%
3 Level
3.8Vrms ±30%
The PURISMA is designed to operate from any 50/60Hz single phase supply 100-240 Vac.
All information on model, serial number, and month/year of manufacture is located on the rear panel.
Environmental Conditions for Transport and Storage
• Temperature -10 to +40
℃
• Relative Humidity 5 to 95%
• Atmospheric Pressure 500 to 1060 hPa
Environmental Conditions for Operation