21
8-4. Actual load calibration (
실부하
교정
)
If the actual load (load, displacement, pressure) is applied to the sensor and there is a
difference between the value displayed in the indicator and the actual load value, the display
value of the indicator can be changed to the actual load value.
Up to 8 points can be stored. As
the calibration interval is differentiated, the linearity of the sensor can be increased.
In the actual
load calibration mode, calibration is possible at the forward (+) output of the sensor and
calibration is not possible at the reverse (-) output of the sensor.
Measuring mode
While
key is being
Press, press
key
Display Function mode
Display Sensor output value calibration mode
Display Actual load calibration mode
After changing to actual load value using
and
key.
Press the
key. (500.0 is an example value arbitrarily
changed)
Actual load is applied to the sensor and the indicated value
flashes.(500.7 is an arbitrary example value)