In case curing needs to be repeated, press the 

Ultraled

 activation button again. In order to 

interrupt the curing cycle, press the activation 
button again.

Bleaching 

Attach the acrylic bleaching tip (fig. B, item 3 or 4) 
to the front orifice of the handpiece. 

 The bleaching tips are sold separately. 
After  the  tip  has  been  attached  to  the 
equipment,  it  should  be  cleaned  with  some 
gauze,  since  oiliness  from  the  user’s  hands 
and other stains on the acrylic tip may affect 
light conduction.   

Plug the equipment in the power source outlet, and 
turn on its master switch.
Follow the gel manufacturer’s instructions as regards 
activation  time  and  the  distance  between  the  tip 
and the gel.
Press the button to activate the LED, which will light 
up  slowly.  As  soon  as  it  is  activated,  the  LED  will 
remain on during the set time and a tone will be 
heard every 10 seconds. 
When the programmed time has ended, the LED will 
go off. (refer to “Time setting”). 

To  interrupt  the  bleaching  cycle,  press  the 
activation button again.

Time Setting 

Activate the master switch to turn on the instrument 
and keep the handpiece activation button pressed 
until  a  long  tone  is  heard.  Release  the  activation 
button. Press it the number of times corresponding 
to the desired tome setting. Each touch corresponds 
to 10 seconds. The maximum setting time is 100 
seconds.

Intensity time adjustment program

Press  the  activation  button  on  the  handpiece 
to  activate  the  LED.  Then,  without  releasing 
the  activation  button,  wait  three  seconds,  and 
a  long  tone  will  be  emitted.  Still  keeping  the 
activation  button  pressed,  wait  until  a  second 
tone is emitted. Release the activation button and 
wait for a tone indicating the first setting phase, 
“Low Power”. Then, press the activation button 
to  set  the  application  percentage.  Each  touch 
corresponds to 10% of the total pre-adjusted total 
functioning time, and the maximum will be 100% 
(in case the activation button is not pressed, the 
system  will  understand  that  no  percentage  has 
been set in this phase).
After the desired application percentage has been 
programmed, and if it is less than 100%, wait 2 
seconds, and 2 tones will be emitted, indicating 

the  second  setting  phase,  “Medium  Power”. 
Repeat the procedure used for the first phase. At 
this moment, only the remaining percentage can 
be  programmed.  “High  Power”  will  correspond 
to  the  remaining  percentage  from  the  two  first 
phases.  Two  seconds  after  the  setting  of  the 
second phase has been concluded, 3 fast tones will 
be emitted, indicating setting completion. In case 
you wish to set only some of the phases, access 
the setting model as explained above, but do not 
press the activation button after each phase. The 
system will automatically shift to the next phase. 
When the activation button is not pressed during 
the setting of phase 1 and 2, the application will 
be set at 100% with “High Power”. 
Example:  For  an  application  time  setting  of  60 
seconds and the phases respectively set at 10, 20 
and 70%, the LED will remain on for 6 seconds at 
low power, for 12 seconds at medium power and 
for 42 seconds at high power.

Cleaning

Cleaning

Cleaning

Clean  the  equipment  surface  with  a  clean  soft 
cloth moistened in Aplic Odonto by Dabi Atlante or 
another product with similar chemical properties

The  use  of  other  chemical  products  is  not 
recommended as it may damage the equipment. 
In case PVC film is used, make sure that the 
ventilation orifices (air inlet and outlet) are not 
clogged. The acrylic light tip should never be 
wrapped in any kind of material.

Sterilization

Sterilization

Sterilization

The acrylic curing light tip as well as the bleaching 
light tips must not be autoclaved. 

Maintenance

Maintenance

Maintenance

Preventive maintenance

In  order  to  reduce  equipment  faults  and  increase 
durability,  call  Dabi  Atlante  Authorized  Technical 
Assistance and set a regular preventive maintenance 
plan.

Corrective maintenance

In case the equipment presents any faults that are 
not listed in “Troubleshooting”, contact Dabi Atlante 
Authorized Technical Assistance.

Do not open the equipment and/or try to repair 
it as such action may aggravate the problem or 
even generate further malfunctioning.

Circuit diagrams, list of parts, components, 
others

In  case  they  are  necessary,  ask  Dabi  Atlante  for 
electric diagrams and/or lists of parts, components 

Содержание Ultraled

Страница 1: ...ess CEP ZIP code Cidade City Estado Country Instala o Installation Representante Dealer Code T cnico Technician Assinatura do Cliente Signature Customer Dabi Atlante Inds M dico Odontol gicas Ltda Av...

Страница 2: ...de Servi os Autorizada Dabi Atlante 05 Simbologia 05 Recomenda es finais 05 Termos de garantia 06 Caracter sticas t cnicas 06 Falhas causas e solu es 06 Apresenta o Apresenta o Apresenta o Prezado Cli...

Страница 3: ...ogramada a porcentagem de aplica o desejada e sendo ela menor que 100 espere 2 segundos e 2 bips ser o emitidos indicando a segunda etapa de programa o M dia Pot ncia repita o procedimento da primeira...

Страница 4: ...sto pode agravar o problema ou at gerar outras falhas Esquemas de circuito listas de pe as componentes e outros Caso seja necess rio solicite Dabi Atlante esquemas el tricos e ou listas de pe as compo...

Страница 5: ...bi Atlante Rede de Servi os Autorizada Dabi Rede de Servi os Autorizada Dabi Atlante Atlante Os servi os de manuten o nos equipamentos Dabi Atlante dever o ser feitos por meio de uma Assist ncia T cni...

Страница 6: ...trica LED queimado Obstru o da ponteira condutora de luz Aguardar a normaliza o ou solicitar o aux lio t cnico Substituir o cabe ote Providenciar substitui o ou limpeza T cnica Dabi Atlante pois possu...

Страница 7: ...leaning 08 Sterilization 08 Maintenance 08 Recommendations 09 Procedures in case equipment functioning is altered 09 Sensitivity to predictable environmental conditions in normal working situations 09...

Страница 8: ...ercentage has been programmed and if it is less than 100 wait 2 seconds and 2 tones will be emitted indicating the second setting phase Medium Power Repeat the procedure used for the first phase At th...

Страница 9: ...as long as the cleaning maintenance transportation and operation recommendations found in this manual are observed Procedures in case of equipment disposal Procedures in case of equipment Procedures...

Страница 10: ...iding tip Awaitforpowersupplynormalization or call for technical assistance Replace the head Replace and clean it Troubleshooting Troubleshooting Troubleshooting Fragile Located on the packaging side...

Страница 11: ...ess CEP ZIP code Cidade City Estado Country Instala o Installation Representante Dealer Code T cnico Technician Assinatura do Cliente Signature Customer Dabi Atlante Inds M dico Odontol gicas Ltda Av...

Отзывы: