32
Radiation bunch characteristics
The pictures below present radiation bunch characteristics with and without the use of extension
whit colimador cone.
The extension whit colimador cone is used to apply the parallelism technique which gives less
distortion to the image generated due to x-ray from film/sensor presented less inclination related
to reference axis.
Ø16mm
62mm
203.2mm
58mm
Eixo de referência
Reference Shaft
Eje de referencia
Diafragma
Diaphragm
Ponto focal
Focal point
Punto focal
Eixo ânodo / cátodo
Ânodo / cátodo shaf
t
Eje ânodo / cátodo
Ponto focal
Focal point
Punto focal
62mm
304mm
Ø16mm
Ø20,5mm
58mm
Eixo de referência
Reference Shaft
Eje de referencia
Diafragma
Diaphragm
Diafragma
Diaphragm
Eixo ânodo / cátodo
Ânodo / cátodo shaf
t
Eje ânodo / cátodo
Radiation bunch X without extension whit colimador cone (203,2mm)
Radiation bunch X with extension whit colimador cone (304mm)
NOTE:
If the extension whit colimador cone is not used, it does not affect the safety level to patient
during use of this device.
OPERATION OF EQUIPMENT
Содержание Spectro 70X Coluna Movel
Страница 1: ......
Страница 15: ...15 Dimensions mm Model Coluna M vel TECHNICAL SPECIFICATIONS...
Страница 16: ...16 Dimensions mm Model Parede TECHNICAL SPECIFICATIONS...
Страница 17: ...17 Dimensions mm Model Pantogr fico Coluna M vel TECHNICAL SPECIFICATIONS...
Страница 18: ...18 Dimensions mm Model Pantogr fico Piso TECHNICAL SPECIFICATIONS...
Страница 19: ...19 Dimensions mm Model Pantogr fico Parede TECHNICAL SPECIFICATIONS...
Страница 22: ...22 Content of accessible and non accessible demarcations 01 02 TECHNICAL SPECIFICATIONS...
Страница 23: ...23 03 04 05 TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations...
Страница 24: ...24 06 07 08 10 09 TECHNICAL SPECIFICATIONS Content of accessible and non accessible demarcations...
Страница 42: ...42...
Страница 43: ...43...
Страница 44: ......