Emissions Test 

IEC 60601 Test Level 

Actual Level 

Power frequency magnetic 
field immunity test following 
IEC 61000-4-8 

50 Hz, 30 A/m 

50 Hz, 30 A/m 

Voltage dips and interruptions 
immunity test following IEC 
61000-4-11 

30% Reduction for 500 
mS at 0 degrees,  
100% Interruption for 10 
mS at 0, 45, 90, 135, 180, 
225, 270,315 degrees,  
100% Interruption at 5000 
mS at 0 degrees 

30% Reduction for 500 mS at 
0 degrees,  
100% Interruption for 10 mS 
at 0, 45, 90, 135, 180, 225, 
270,315 degrees,  
100% Interruption at 5000 
mS at 0 degrees 

 

Table 5 Electromagnetic Immunity

 

The device is intended for use in the electromagnetic environment specified below. The 
customer or the user of the device should assure that it is used in such an environment. 

Immunity test 

IEC 60601 
test level 

Compliance 
level 

Electromagnetic environment - guidance 

Conducted RF 
IEC 61000-4-6 

 

3 Vrms  
150 kHz to 
80 MHz 

3 Vrms 

 

Portable and mobile RF communications 
equipment should be used no closer to any 
part of the device, including cables, than the 
recommended separation distance calculated 
from the equation applicable to the frequency 
of the transmitter. 

Recommended separation distance 

𝑑 = 1.2√𝑃

 

𝑑 = 1.2√𝑃

 80 MHZ to 800 MHz 

𝑑 = 2.3√𝑃

 800 MHz to 2.5 GHz 

 
where 

P

 is the maximum output power rating 

of the transmitter in watts (W) according to 
the transmitter manufacturer and

 d

 is the 

recommended separation distance in meters 
(m). 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey, 
should be less than the compliance level in 
each frequency range. 

a

 

Interference may occur in the vicinity of 
equipment marked with the following symbol: 

Radiated RF 
IEC 61000-4-3 

3 V/m 
80 MHz to 
2.5 GHz 

3 V/m 

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people 

Содержание 1 Series

Страница 1: ...MNL 00049 1 Proprietary and Confidential Page 1 of 19 DT Home Control Operation Manual Modes of Operation Timed Mode...

Страница 2: ...10 3 How to Position Yourself 10 11 0 Setting Up a Treatment Time 10 12 0 Viewing Device Code for Flex Rx 11 13 0 Enable Disable Refill FlexRX 11 13 1 Refill FlexRx 12 14 0 Special Notes 12 15 0 Care...

Страница 3: ...have prescribed your device with FlexRx a system that limits the total number of exposures the device will deliver Specific information about these systems is provided in this manual 2 0 Indications...

Страница 4: ...vice department to determine an appropriate replacement 5 0 Unpacking and Assembly 1 Open the top of the box and grasp both sides of the device Pull it out of the box along with the foam packers on ea...

Страница 5: ...kin that the physician does not intend to treat During extremely long treatments some surfaces may become hot Remove any body parts from those surfaces if it becomes uncomfortable The unit should have...

Страница 6: ...0 Environmental Specifications If the device continues to malfunction cease operation and contact the Daavlin Service Department If necessary operation of equipment may be terminated by disconnecting...

Страница 7: ...ee page 21 DT Controlled Phototherapy Devices are considered CISPR Group 1 Class A devices to be used for prescription use only 8 0 Operating Specifications Ambient Temperature 15 C to 30 C 59 F to 86...

Страница 8: ...a prominent and easily readable position Please read the label carefully as it contains important safety information Figure 2 Warning Label Figure 3 Identification Label The following table lists all...

Страница 9: ...treatments along with any other notes regarding your treatment or treatment results For example forgot lip balm put sunscreen on tender area etc Note A free printable treatment log is available for d...

Страница 10: ...e on the acrylic surface Excess pressure can cause the acrylic surface to crack and break Larger Area Treatment Choose the 9 inches position when starting your therapy Position the 1 Series so that th...

Страница 11: ...n again will pause treatment 8 While paused pressing the CANCEL button will end the treatment and prompt a new treatment 12 0 Viewing Device Code for Flex Rx The Device Code is predetermined by the so...

Страница 12: ...ed When power is restored the controller will prompt for the Key Code and another treatment will have to be started a If Flex Rx is activated and the controller loses power during a treatment an expos...

Страница 13: ...leaners 2 Allow to air dry and inspect for visible contaminants 3 If contaminants remain repeat until no visible containments remain repeat steps 0 and 2 15 2 3 High Level Disinfection Follow a high l...

Страница 14: ...removed 4 On top of each lamp socket there will be a red button that when pushed will allow you to remove the base of the lamp from its corresponding lamp socket With one hand firmly press and hold th...

Страница 15: ...tual Level Electrostatic discharge immunity test following IEC 61000 4 2 8kV conductive surfaces coupling planes 2kV 4kV 8kV and 15kV non conductive surfaces 8kV conductive surfaces coupling planes 2k...

Страница 16: ...guidance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms Portable and mobile RF communications equipment should be used no closer to any part of the device including cables than the recomm...

Страница 17: ...elephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF tran...

Страница 18: ...with our operating instructions The Limited Warranty does not apply to Equipment failure or damage caused by shipping or customer abuse misuse or accident incorrect customer installation improper ele...

Страница 19: ...d warranty service work However this Limited Warranty will be declared null and void if non qualified technicians perform any repair or maintenance on the Equipment unless prior written authorization...

Отзывы: