background image

 

Detector Number

Gammas counted by 

indicted Detector

Neutrons counted by 

indicated Detector

Detector Number

Gamma Alarm

Level

Neutron Alarm

Manual Alarm 

Reset Button

Detector Number

Gammas counted by 

indicted Detector

Neutrons counted by 

indicated Detector

Detector Number

Gamma Alarm

Level

Neutron Alarm

Manual Alarm 

Reset Button

System Operation  

 
Non-Alarming  
 
When not alarming, the rad-D information panel displays the measured gamma counts per second and 
measured neutrons per second for each detector in a repeating display cycle. If your rad-D system 
consists of four gamma detectors, the display would show the gammas/second from detector #1, 
indicated by the 1

st

 

 

LED light below the gamma count display being lit. Then the gammas/second from 

detector #2 would be shown (accompanied by the second LED light being lit). The cycle would continue 
with the gammas/second measured by detectors #3 and #4 being shown and then would begin again with 
detector#1.  A similar cycle would be shown if there were four neutron detectors. If you have a mixed 
combination of gammas and neutron detectors, than the detectors would be cycled similarly to the 
previous examples, except split between the gamma and neutron displays.  
 
Alarms  
 
During an alarm, the non-alarming detectors are no longer displayed. Instead the gamma or neutron 
count of the alarming unit is displayed, with the specific detector identified by the detector number light. If 
more than one gamma or neutron detector is alarming, the display shifts between the two. Both the 
gamma and neutron counter displays are real-time. When the radiation source moves on, they will go 
back to displaying normal levels. The Alarm Display Panel also has persistent alarm displays for gamma 
and neutron alarms. These will not clear to normal levels until reset, either using the manual button on the 
Alarm Display or via computer. The persistent gamma alarm displays the highest alarm level (1 through 9) 
that a detector measured. The persistent neutron alarm indicates an alarm, but does not display a level 
(any neutron source above natural background requires investigation). 
 

Содержание rad-D

Страница 1: ...Users Manual LAURUS Systems Inc Ph 410 465 5558 Fax 410 465 5257 www LaurusSystems com...

Страница 2: ...stem Components A rad D system consists of one to four detectors Figure 1 an Alarm Display Panel Figure 2 and optional PRM Mounting Bracket Figure 3 There are two types of detectors a gamma detector a...

Страница 3: ...rface the receiving computer must be within 100 meters or less depending on intervening walls of the Alarm Display Panel If the rad D is integrated as part of a computer network via its Serial Port in...

Страница 4: ...ter that can be detected by the gamma detector to avoid false alarms please do not aim the front of the detector into the X ray machine Either aim the gamma detector at the conveyor belt coming into t...

Страница 5: ...r 1 A similar cycle would be shown if there were four neutron detectors If you have a mixed combination of gammas and neutron detectors than the detectors would be cycled similarly to the previous exa...

Страница 6: ...rad D component while it is still plugged into a power source Please use proper care and good practices when lifting heavy rad D components Notification This equipment has been tested and found to com...

Страница 7: ...rant any installation maintenance or Service that it did not perform SERVICE WORK PERFORMED BY SERVICE CENTERS NOT AUTHORIZED BY D TECT SYSTEMS TO PERFORM SUCH WORK WILL VOID THIS WARRANTY What This W...

Страница 8: ...grant you a credit for such Products or parts as depreciated and accept their return The depreciation will be an equal amount per year over the lifetime of the Products accessories battery or parts a...

Отзывы: