6
DCS-9500T Quick Installation Guide
ENGLISH
Temperature Management
1.
Navigate to
Configuration
>
Temperature Measurement
.
2.
There are 3 main configuration sections that need to be modified; Parameter Configure, Thermal
Mapping, and Thermal Calibration.
Thermal Calibration
Note
: It necessary to configure these settings when using the Blackbody.
1.
Set Enable and Display Area Info to
On
to enable Thermal Calibration.
2.
The Target Temperature should be set to the LED value displayed on the back of the Blackbody as
seen above. 40 °C is recommended.
3.
Emission Rate is the thermal radiation emissivity of the Blackbody. The default value is 0.98.
4.
Distance is the distance from the Blackbody to the thermal camera.
5.
Using the mouse, drag a square over where the Blackbody is located on the screen, as seen in the
above image.
Note
: Be sure that the green square exactly outlines the Blackbody for the most
accurate results.
6.
Click
Apply
to save settings.