18
* Fractional pre-vacuum procedure = steam sterilization with repetitive pre-vacuum. The procedure used here is to
perform steam sterilization through three pre-vacuums.
4.9 Storage
1. Store in a clean, dry, ventilated, non-corrosive atmosphere with a relative humidity of 10% to 93%, an atmospheric
pressure of 70KPa to 106KPa, and a temperature of -20 ºC to +55 ºC;
2. After sterilization,the product should be packaged in a medical sterilization bag or a clean sealing container, and
stored in a special storage cabinet. The storage time should not exceed 7 days. If it is exceeded, it should be repro-
cessed before use.
Notes:
a) The storage environment should be clean and must be disinfected regularly;
b) Product storage must be batched and marked and recorded.
4.10 Transportation
1. Prevent excessive shock and vibration during transportation, and handle with care;
2. It should not be mixed with dangerous goods during transportation.
3. Avoid exposure to sun or rain or snow during transportation.
The cleaning and disinfection of main unit are as follows.
• Before each use, wipe the surface of the main unit, the connector of cable and cable with a soft cloth or paper towel
soaked in 75% medical alcohol. Repeat the wipe for at least 3 times.
• Before each use, please let the Ultrasonic scaler works for 20-30 seconds at maximum water volume, then install the
handpiece.
• After each use, please let the Ultrasonic scaler works for 20-30 seconds at maximum water volume, then remove the
handpiece.
• After each use, wipe the surface of the main unit, the connector of cable and cable with a soft cloth soaked in clean
water (distilled or deionized water) or a clean disposable wipe. Repeat the wipe for at least 3 times.
Содержание D SCADLER U
Страница 1: ...EN_ User manual ES_ Manual de instrucciones D_SCADLER U...
Страница 6: ...6 1 4 4 List of removable parts picture 4 picture 5 picture 6 picture 5...
Страница 23: ...ENGLISH 23...
Страница 24: ...24...
Страница 26: ......
Страница 30: ...30 D_Scaler Led U Punta Indicador Endo Indicador Perio...
Страница 32: ...32 1 4 4 Lista de partes removibles...
Страница 48: ...48 7 Instrucci n de s mbolos...
Страница 50: ...50...
Страница 51: ...ESPA OL 51...
Страница 53: ...ZMN SM 128 ES V1 0 20220307 PROCLINIC S A U C Palermo 9 50197 Zaragoza Espa a Rev 11 22 0197...