JF-1A Conductivity Sensor
A440-010
December, 2005
19
TERMINAL Function
1
Data/Power Ground
2
RS-232 Data In (RXD)
3
Power Input +
4
RS-232 Data Out (TXD)
The RS-232 Interface is fully isolated from the sensor electronics through the
use of optical isolators located in the JF-1A/ATEX sensor.
Data cable for the RS-232 should be low capacitance data cable. Lengths in
excess of 20 meters may result data degradation of the RS-232 signals.
4.9 Power Input (4-Wire)
WARNING
: Power supply must be fully isolated and meet all local and site
codes for its use.
Power input to the 4-Wire power connections should be clean filtered DC
voltage in the range of 7 – 35 VDC. Current consumption is approximately 10
mA.