![CYRUSTEK ES269 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/cyrustek/es269/es269_manual_2706890026.webp)
Ver 1.2
26
22/02/24
ES269
6000 Counts Intelligent Auto-Function DMM IC
Main_Digit1 –Main_Digit3
Main_Digit1 – Main_Digit3 is the readings of measurement result shown on LCD panel.
M_DigitN
3
– M_DigitN
0
consist of 4-bit BCD code. The M_SIGN is the sign bit of readings.
D0
D1
D2
D3
D4
D5
D6
D7
a04
M_SIGN
X
X
X
X
X
X
X
a05
M_Digit3
3
M_Digit3
2
M_Digit3
1
M_Digit3
0
M_Digit2
3
M_Digit2
2
M_Digit2
1
M_Digit2
0
a06
M_Digit1
3
M_Digit1
2
M_Digit1
1
M_Digit1
0
M_Digit0
3
M_Digit0
2
M_Digit0
1
M_Digit0
0
FUNCTION
The packets of
a07-a09
indicate the measurement mode of the meter. The following table
summarizes the transmitted bit for each mode.
For example, if the meter operates in Voltage mode, this
a07
packet is
10000000
.
D0
D1
D2
D3
D4
D5
D6
D7
a07
V
A
Ohm
Continuity
Diode
Capacitance
Hz
X
a08
X
X
X
X
EFmode
SCAN
X
X
a09
X
X
X
X
X
X
X
X
STATUS
The a10-a13 packets indicate the whole status when ES258 is in normal operation. For example,
if meter is operated at ACV / MANU range, then
a10
packet will set
011000xx
.
The format of the four packets is shown below.
D0
D1
D2
D3
D4
D5
D6
D7
a10
AUTO
MANU
AC
DC
OL
X
1
APO
a11
X
X
LBAT
LBAT30s
X
X
X
X
a12
X
X
X
X
X
X
MAX
MIN
a13
Danger
OPEN
SHORT
DISCH
X
X
X
X
AUTO
: When auto range is selected.
MANU
: When manual mode is selected.
AC
: When AC mode is selected.
DC
: When DC mode is selected.
OL
: When ‘OL’ is displayed on LCD.
APO
: When APO is displayed on LCD.
LBAT
: When battery voltage is too low.
LBAT30s
: When ‘Lobt’ is shown on LCD.