![CYRUSTEK ES269 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/cyrustek/es269/es269_manual_2706890010.webp)
Ver 1.2
10
22/02/24
ES269
6000 Counts Intelligent Auto-Function DMM IC
VAHz function
ES269 can automatically measure frequency of ACV or ACA and show the value on sub-display. The
maximum frequency range is 60KHz. For correct readings, the signal input level must exceed 10% of
full scale. The reading value will be shown on the sub-display in voltage or current mode.
Table 8. frequency range
Mode
Full Scale
FR1
600.0Hz
FR2
6.000KHz
FR3
60.00KHz
Resistance Measurement
A divider automatically provides a suitable full-scale range in resistance measurement mode. The
following table summarizes the full-scale ranges and the reference resistors in each configuration.
In scan mode, it automatically switches to resistance measurement as the resistance value exceed 600.1
Ω
.
In auto mode, ES269 can automatically switch to appropriate range for measurement.
In manual mode, User can switch range by Range key.
Table 9. Resistance ranges
Configuration
Full Scale Range
Reference Resistor
Equivalent value
R1
600.0Ω
3
OR1
200Ω
R2
6.000KΩ
VR4
2KΩ
R3
60.00KΩ
VR3 || VR1
20KΩ
R4
600.0KΩ
VR2 || VR1
200KΩ
R5
6.000MΩ
VR1
2MΩ
R6
60.00MΩ
4
OR2
20MΩ
3
continuity check is implemented in this range
4
60MΩ range is only available in non-scan mode.