M5-CYG4P-M-01_Iss5.doc
Cygnus 4+ Operating Manual
47
1.
Access the Menu and scroll down
to the ‘Calibration’ group. Then
scroll right and select ‘Velocity’.
Press the Ok/Select key to open
the ‘Velocity’’ function
3.
If there has been a previous
Calibration a message will warn
that changing the velocity will
alter the calibration;
Press the OK/Select key to
continue, or X/Cancel to abort.
4.
Use the Up and Down keys to
adjust the velocity value as
required
Use the Left and Right keys to
highlight the digit to change.
5.
When done press the OK/Select
key to save the changes.
Or press the X/Cancel key to exit
without saving.
When the velocity has been manually set the value is displayed at
the top of the measurement screen in the status area.
Velocity of Sound value ‘5920 m/s’