background image

Shenzhen Zhiyong Electronics Co., Ltd.

 

www.cybertek.cn 

High Voltage Differential Probes 

DP6000 Series 

DP6070   700Vpk/ 70MHz 

DP6150  1500Vpk/ 70MHz 

DP6150A 1500Vpk/100MHz 

DP6150B 1500Vpk/200MHz 

DP6280  2800Vpk/100MHz 

DP6700  7000Vpk/ 70MHz 

DP6700A 7000Vpk/100MHz 

ShenZhen ZhiYong Electronic Co.,Ltd 

CYBERTEK

Содержание DP6000 Series

Страница 1: ...k cn High Voltage Differential Probes DP6000 Series DP6070 700Vpk 70MHz DP6150 1500Vpk 70MHz DP6150A 1500Vpk 100MHz DP6150B 1500Vpk 200MHz DP6280 2800Vpk 100MHz DP6700 7000Vpk 70MHz DP6700A 7000Vpk 100MHz ShenZhen ZhiYong Electronic Co Ltd CYBERTEK ...

Страница 2: ...er the measurement and then remove the probe When BNC lines are connected to the oscilloscope or other devices ensure the BNC terminal is well grounded Please check the probe skin If there is any breakage stop using it immediately Select the product standard adapter power supply DP6000 Series Brief Description Modal Maximum Input Differential Voltage Bandwidth Attenuation DP6070 700V 70MHz 10X 100...

Страница 3: ...part of probe 4 Accessories 5 Specifications 7 Machinery specification 8 Environment Specifications 9 Operating steps 9 Test mode Offset Settings 9 Precautions For Use 10 Performance Verification 11 Installation 12 DC accuracy 12 Rise time 12 DC common mode rejection ratio CMRR 13 Test records form 14 Maintenance 14 Warranty 14 Packing List 15 CYBERTEK ...

Страница 4: ...larming function which can also be turned off manually It can be powered by USB port which is convenient for the customers The probe is standard BNC output port which can be connected to oscilloscope of any brand The probes have great CMRR which is very important in many power electronics applications 2 Application Floating voltage measurement Inverter Switch Power Supply Welding plating power sup...

Страница 5: ...t attenuation indicates different ranges such as DP6150A 500X indicate the maximum test voltage is 1500V 50X presents maximum test voltage is 150V DP6700A 1000X shows maximum test voltage is 7000V 100X indicate maximum test voltage is 700V oscilloscope attenuation factor should be set accordingly based on the probe attenuation selection BANDWIDTH The series products have bandwidth selection functi...

Страница 6: ...r interface Standard USB type B interface supply power with standard USB adapter can be supplied by oscilloscope easy to use also can be supplied by portable power source convenient for outdoor test Factory Setting The default factory setting is high attenuation ratio FULL bandwidth audible alarm is on The product has automatic memory automatically save the state before power off Accessories Descr...

Страница 7: ...1000V CATIV 600V Alligator Clips CK 262 CATIII 1000V CATIV 600V Pincer Clips CK 281 CATIII 1000V Hook Clips CK 284 CATII 1000V Extender Leads CK 301 CATIII 1000V BNC Output Line CK 310 Double ended BNC connector coaxial line 1m USB Line CK 315 1 5m Power Adapter CK 605 USB 5V 1A NOTE The above refers to non standard accessory of this model CYBERTEK ...

Страница 8: ...00V CATIII 1000V CATII 1000V CATIII 2300V CATI Input impedance Single ended to ground 2 5MΩ 5MΩ 5MΩ 20MΩ Between inputs 5MΩ 10MΩ 10MΩ 40MΩ Input capacitance Single ended to ground 4pF 4pF 4pF 5pF Between inputs 2pF 2pF 2pF 2 5pF CMRR DC 80dB 80dB 80dB 80dB 100kHz 60dB 60dB 60dB 60dB 1MHz 50dB 50dB 50dB 50dB Noise Vrms 10X 20mV 50X 50mV 100X 100mV 100X 200mV 100X 120mV 500X 300mV 1000X 600mV 1000X ...

Страница 9: ...Frequency 5 Mechanical Specification Model Parameters Input leads Approx 28cm 17cm for DP6150B Extender leads CK 301 Approx 1m BNC Output Line CK 310 Approx 1m Alligator clips CK 261 Approx 85 40 17mm Alligator clips CK 262 Approx 106 43 16mm Pincer clips CK 281 Approx 152 50 13mm Hook clips CK 284 Approx 121 23 23mm Probe dimensions Approx 195 65 28mm Probe weight DP6070 Approx 216g DP6150 A Appr...

Страница 10: ... the probe clips based on needs start after connecting to the circuits to be tested When testing the probe body should keep away from high voltage pulse circuits to reduce interference to the probe Turn off the probe power after the testing is completed first disconnect the two inputs from the tested points and then unplug the BNC plug from the oscilloscope 8 Test Mode Offset Setting User may ente...

Страница 11: ... frequency response Please view below for the winding method It is better not to extend input lead when testing otherwise it may introduce more noise If extra extension lead is necessary please ensure the extension leads are at same length and the input frequency is under 5MHz errors may exist if exceeds 5MHz output Without extender leads With extender leads Note CYBERTEK ...

Страница 12: ...O4000 Displays probe output Standard signal generator calibrator Amplitude accuracy 0 75 rise time 3ns e g FLUKE WAVETEK 9100 Test bandwidth AC accuracy common mode rejection ration Digital multimeter Accuracy of not less than 6 and a half e g KEITHLEY 2000 Test the DC accuracy Insulation pincer clips Supplied in the accessories Testing clips BNC adapter 1 BNC male to female dual show as Figure 1 ...

Страница 13: ... Repeat step 4 6 and calculate whether is within the accuracy ranges Model Attenuation Rate Signal source output voltage Probe expected output voltage Probe practical output voltage DP6070 10X 1V 100mV 2mV 100X 10V 100mV 2mV DP6150 A B 50X 5V 100mV 2mV 500X 50V 100mV 2mV DP6280 100X 10V 100mV 2mV 1000X 100V 100mV 2mV DP6700 A 100X 10V 100mV 2mV 1000X 100V 100mV 2mV 10 3 Rise Time Configure the fas...

Страница 14: ...0X 50X 100X Set 500V DC voltage for signal source now the voltage output shut up Connect the two probe inputs to 500V voltage Connects the probe output to BNC female to dual binding post as shown in Figure 3 and plug into the two inputs of the digital multimeter Enable signal source output respectively record voltage output values check with the following chart to calculate whether is within the r...

Страница 15: ... 5ns DP6150B 50X 1 75ns 500X 1 75ns DP6280 100X 3 5ns 1000X 3 5ns DP6700 100X 7ns 1000X 7ns DP6700A 100X 3 5ns 1000X 3 5ns DC common mode rejection ratio DP6070 10X 1mV DP6150 A 50X 1mV DP6280 100X 1mV DP6700 A 100X 1mV 11 Care and Maintenance Keep the probe clean and dry Please wipe with soft dry cloth when clean needed must not use chemicals to clean Please put the probe in the package provided ...

Страница 16: ...pcs 1 pcs 1 pcs 1 pcs USB connecting line AM BM 1 5m 1 pcs 1 pcs 1 pcs 1 pcs Advanced tool cabinet 1 unit 1 unit 1 unit 1 unit User manual 1 book 1 book 1 book 1 book Warranty card 1 unit 1 unit 1 unit 1 unit Testing report 1 pcs 1 pcs 1 pcs 1 pcs NOTE The above refers to non standard accessory of this model CYBERTEK SHENZHEN ZHIYONG ELECTRONICS CO LTD Addr Room A1702 Building 4 TianAn Cyber Park ...