+5V
+GND
Port A0
Figure 3-9. Schematic showing switch detection by digital channel Port A0
For more information on digital signal connections
For more information on digital signal connections and digital I/O techniques,
available on the software CD
which accompanies your device.
Power terminals
The
PC +5V
connection (pin 30) draws power from the USB connector. This terminal is a 5 V output that is
supplied by the host computer. Refer to the "
" diagrams on page 3-3 for the location
of this pin.
Caution!
The +5 V terminal is an output. Do not connect to an external power supply or you may damage
the
UMDAS 0802DA
and possibly the computer.
The maximum total output current that can be drawn from all
UMDAS 0802DA
connections (power, analog and
digital outputs) is 420 mA. This maximum applies to most personal computers and self-powered USB hubs.
Bus-powered hubs and notebook computers may limit the maximum available output current to 100 mA.
Just connecting the
UMDAS 0802DA
to your computer draws 80 mA of current from the USB +5 V supply. Once
you start running applications with the
UMDAS 0802DA
, each DIO bit can draw up to 2.5 mA, and each analog
output can draw 15 mA. The maximum amount of +5 V current available for experimental use, over and above
that required by the
UMDAS 0802DA
, is the difference between the
total
current
requirement
of the USB (based on
the application), and the
allowed
current
draw
of the PC platform (500 mA for desktop PCs and self-powered
hubs, or 100 mA for bus-powered hubs and notebook computers). With all outputs at their maximum output
current, you can calculate the total current requirement of the
UMDAS 0802DA
USB +5 V supply as follows:
(
UMDAS 0802DA
@ 80 mA) + (16 DIO @ 2.5 mA ea) + (2 AO @ 15 mA ea ) = 150 mA
For an application running on a PC or powered hub, the maximum available excess current is
500 mA
−
150 mA = 350 mA. This number is the total maximum available current at the PC +5 V screw
terminals.
CyberResearch, Inc.
highly recommends that you figure in a safety factor of 20% below this
maximum current loading for your applications. A conservative, safe user maximum in this case would be
280 mA.
Since laptop computers typically allow up to 100 mA, the
UMDAS 0802DA
in a fully-loaded configuration may be
above that allowed by the computer. In this case, you must determine the per-pin loading in the application to
ensure that the maximum loading criteria is met. The per-pin loading is calculated by simply dividing the +5 V
by the load impedance of the pin in question.
External trigger terminal
The
TRIG_IN
connection (pin 18) can be configured for either rising or falling edge. Refer to the "
" diagrams on page 3-3 for the location of this pin.
Counter terminal
The
CTR
connection (pin 20) is input to the 32-bit external event. Refer to the "
diagrams on page 3-3 for the location of this pin. The internal counter increments when the TTL levels
transition from low to high. The counter can count frequencies of up to 1 MHz.
3-7
UMDAS 0802DA
User's Guid
e F
unctional Details
Содержание UMDAS 0802DA
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...UMDAS 0802DA CyberResearch Data Acquisition iv Copyright 2006 CyberResearch Inc Intentionally Blank...
Страница 35: ......
Страница 36: ......
Страница 38: ...UMDAS 0802DA CyberResearch Data Acquisition 32 Copyright 2006 CyberResearch Inc Intentionally Blank...
Страница 40: ...UMDAS 0802DA CyberResearch Data Acquisition 34 Copyright 2006 CyberResearch Inc Intentionally Blank...
Страница 41: ......