- 34 -
C
leaninG
anD
DisinfeCtion
of
s
P
o2
s
oft
t
iP
sensor
Aids for cleaning / disinfection:
• Approved cleaning and disinfecting agent
without protein‑fixing effect (always observe the
recommendations of the manufacturer when mixing).
• Compressed air
• Soft, disposable cloths
• Automated cleaning brushes
• Demineralised water
Commercially available cleaning and disinfecting agents
approved for this purpose and based on aldehydes, alco‑
hols, amines, bases or quaternary ammonium compounds
are suitable for cleaning and disinfecting the SpO2 sensors
as long as they exhibit similar active ingredient composi‑
tions and concentrations to the examples listed below.
m
anual
Clean
i
nG
,
Des
i
nfeCt
i
on
of
s
oft
ti
P
s
ensor
1. The sensor must be carefully cleaned or disinfected
after every use, and before use with a different pa‑
tient.
It is recommended that sensors be prepared for reuse
immediately after use since dried residue can be
difficult to remove. Avoid overfilling instrument sieves
and wash trays.
3 Disconnect the sensor from the monitor before cle‑
aning or disinfecting.
4. Thoroughly rinse the sensor to remove surface dirt
Productname
Manufacturer
use
Gigasept Instru AF (Glycoderivat)
Schülke & Mayr GmbH
www.schuelke‑mayr.com
Manual cleaning and disinfection
Gigasept FF (Aldehyd)
Manual cleaning and disinfection
Perfektan TB
Dr. Schumacher GmbH
www.schumacher‑online.com
Manual disinfection
Descoton forte (Aldehyd)
Manual disinfection
Neodisher LM2
Chem. Fabrik Dr. Weigert GmbH & Co. KG
www.drweigert.de
Manual cleaning
t
ested
alKal
i
ne
clean
i
ng
and
d
i
s
i
nfect
i
ng
agents