On the ec3000e, you adjust the saddle height manu‑
ally with a clamping lever. Optional, the saddle height is
electrically adjusted with the corresponding keys on the
control terminal (the display below the saddle indicates the
saddle height).
With the pedal in the bottom position, the angle between
the axis formed by the upper body and the thigh should be
approximately 10°.
Set the handlebar to a position where the patient sitting
upright on the saddle can reach it easily.
To do so, open clamping lever 1 and set the handlebar to a
suitable angle.
On the ec3000e 200, the height of the handlebar can ad‑
ditionally be adjusted with clamping lever 2 ‑ the horizon‑
tal bar should be approximately at the samel level as the
saddle.
a
djusting
saddle
and
handlebar
1 Adjustment of the handlebar angle
2 Adjustment of the handlebar height
(ec3000e 200 only)
3 Adjustment of the saddle height (ec3000e 100 only)
4 Saddle height display (ec3000e 200 only)
Note
• Tighten the clamping levers only as far as necessary, NOT
with maximum force.
• Lubricate the clamping lever threads quarterly at mini-
mum, using a suitable lubricant (e.g. OKS470).