PIM 417-O
PITOT-STATIC LEAK TESTER TYPE D51600
Issue 5
November 2009
Page 29
H. Press the SEL key to move to the next Set Point for calibration (-170 mbar).
J. Set the Pressure Controller to generate a pressure of -170 mbar. Allow the Controller to stabilise at
the control point.
K. Press the START key on the PSLT to commence calibration. If required, note the initial PSLT
reading, e.g. 'Currently -172.4 mB' as shown in
Figure 23
.
Figure 23 Calibrating at Second Set Point
L. Repeat the procedure for each of the remaining Set Points.
NOTE:
Whilst in this menu, the SEL key may be used to skip through the different Set
Points to reach a particular Set Point. This facility is useful, for example, if the
Operator is unhappy with the calibration of any Set Point and wishes to repeat
it without repeating the complete sequence. However, for full calibration, all Set
Points must be individually calibrated.
M. Press the STOP key to exit the Calibration routine at any stage and return to the Maintenance
Menu 3 screen.
Following Calibration, the equipment is subjected to a functional check to confirm that calibration has
been carried out correctly.
4. POST-CALIBRATION FUNCTIONAL CHECKS
A. Select Maintenance Menu 2, PRESSURE MONITOR.
NOTE:
Within this menu, pressure is monitored in mbar only.
B. Set the Pressure Controller to generate a pressure of -340 mbar. When pressure has been
achieved, allow the' Controller to settle at the control point.
C. Confirm that the pressure recorded for the PSLT is within ± 1 mbar.
D. Set the Pressure Controller to generate each of the remaining pressures in turn and confirm that
the PSLT readings are again within ± 1 mbar.
NOTE: Tolerances apply only to measurements at ambient temperature.
E. Set the Controller to generate POTD. When the PSLT is vented, press the PWR key to power
down the equipment and disconnect it from the Pressure Controller. Replace the blanking cover
over the PSLT port to protect it from ingress of contaminants.
Содержание D51600
Страница 1: ...PITOT STATIC LEAK TESTER TYPE D51600 Operation Calibration Manual PIM 417 O Issue 5 November 2009 ...
Страница 2: ......
Страница 11: ...PIM 417 O PITOT STATIC LEAK TESTER TYPE D51600 Issue 5 November 2009 Page 5 Figure 1 Micro Leal Case Assembly ...
Страница 12: ...PIM 417 O PITOT STATIC LEAK TESTER TYPE D51600 Issue 5 November 2009 Page 6 THIS PAGE IS INTENTIONALLY BLANK ...
Страница 16: ...PIM 417 O PITOT STATIC LEAK TESTER TYPE D51600 Issue 5 November 2009 Page 10 THIS PAGE IS INTENTIONALLY BLANK ...
Страница 36: ...PIM 417 O PITOT STATIC LEAK TESTER TYPE D51600 Issue 5 November 2009 Page 30 THIS PAGE IS INTENTIONALLY BLANK ...
Страница 37: ......
Страница 39: ......