TENSPros
Technical specifications for Electrical Muscle Stimulation
(EMS) mode
Waveform:
Mono-phase square pulse wave
Pulse amplitude
Adjustable, 0~105mA peak at 1000 ohm Load each
channel, 1mA/Step.
Pulse Width
Adjustable, from 50 to 300
μ
S microseconds,
Pulse Rate
Adjustable, from 1 to 150 Hz, 1 Hz/step
Contraction time
Adjustable, 1~60 seconds , 1 Sec./ step
Relaxation (OFF) time Adjustable, 0~60 seconds , 1 Sec./ step
Ramp time
Adjustable, 1~6 seconds, 1 Sec. / step, The “On” time
will increase and decrease in the setting value.
Synchronous (S)
Stimulation of both channels occurs synchronously.
The “ON” time including “Contraction”, “Ramp Up” and
“Ramp Down” time. ON TIME=Contr Ramp up
+ Ramp down
Alternate (A)
The Stimulation of the CH2 will occur after the 1st
operation of CH1 is completed. In this program, The
“ON” time including “Contraction”, “Ramp Up” and
“Ramp Down” time. The OFF Time should be equal or
more than the ON Time
ON TIME=Contr Ramp up + Ramp down
OFF TIME
≥
ON TIME
Delay (D)
The Stimulation of the CH2 will occur after CH1 is
s Delay Time. In this program, The “ON” time
includes “Contraction”, “Ramp Up” and “Ramp Down”
time. ON TIME = Contr Ramp up + Ramp
down time. The OFF Time should be equal or more
than the ON Time + Delay Time. Time Delay time is
adjustable from 1 to 10 seconds in this program.
Содержание InTENSity Twin Stim III
Страница 1: ...T E N S P r o s TWIN STIM III INSTRUCTION MANUAL ...
Страница 7: ...T E N S P r o s 1 4 Contraindications 1 5 Warnings Cautions and Adverse Reactions 7 ...
Страница 8: ...T E N S P r o s 8 ...
Страница 9: ...T E N S P r o s out of reach of children you are in any doubt whatsoever CAUTIONS ...
Страница 11: ...T E N S P r o s Adverse Reactions ...
Страница 12: ...T E N S P r o s 2 Presentation 2 1 Front and Rear Panel 12 ...
Страница 37: ...T E N S P r o s ...
Страница 38: ...T E N S P r o s ...
Страница 39: ...T E N S P r o s ...