Table 2: Troubleshooting 

Fault

Possible Cause

Corrective Action

Panel meter fails to 

indicate a reading.

Power switch is off.

Turn switch on.

Power cord not plugged in.

Plug in power cord

Fuse blown on rear panel 
of indicator.

Replace the fuse.

No power coming from power 
source.

Check the service fuses, cir-
cuit breakers, and wiring 
associated with power source 
and repair as needed.

Defective interconnecting cable.

Check continuity and replace. 
if needed.

CAUTION

When checking diode or connec-
tions to diode, do not use a multi
meter which could subject the diode 
to more than five milliamperes for-
ward current, or more than 200 
volts reverse bias.  Excess current 
or voltage will permanently damage 
the diode.

Connections to the diode sensor are 
loose or disconnected.

Check the continuity at the 
cryopump connections pins 3 
and 4.  Repairs must only be 
made by a qualified 
technician.

Polarity of diode connection is 
incorrect.

Check the polarity.

Defective front panel meter.

Check the millivolt output of 
electronics.  It should be 0-35 
mv.  Replace the meter if the 
signal is present and the 
meter is not reponding.

Defective electronics.

Repair as required by a    
qualified technician.