8040024 Rev. A (6/94)
2
Specifications
Electrical
103 - 126V, Single Phase, 50/60Hz or
187 - 253V, Single Phase, 50/60Hz
6 - foot input power cable
0.1 amp fuse on rear panel
10.0 - microampere constant current output to diode
System Accuracy
10 - 27K:
± 2.5K
+2.5% of meter reading
27 - 80K:
± 3.5K
-2.5%, +2.5% of meter reading
80 - 90K:
± 4.5K
-2.5%, +2.5% of meter reading
90 - 100K:
± 5.5K
-2.5%, +2.5% of meter reading
100 - 320K:
± 2.5K
+2.5% of meter reading
Readout Meter
Analog panel meter with temperature scale in Kelvin range 10K to 320K (100
microampere full scale).
Setpoints
2 set points adjustable over 10K to 320K range. Screwdriver adjustable through holes
in top panel. Relays - SPDT with 1.0amp contact. Both NC and NO contacts on
terminal board on rear panel.
Shipping Weight
5 pounds
Installation
1. Connect the cryopump to the temperature indicator SENSOR connector using the
10 - foot interconnecting cable provided with the indicator. Contact
CTI-CRYOGENICS if the supplied cable is too short to order a longer cable.
2. Insert input power cable plug in appropriate power receptacle. The unit is
furnished with either 115V or 208/230V three prong as specified in the sales
order.