background image

Assist

X Mobil

INSTRUCTIONS

FOR USE

US/CA/AU

Содержание AssistX Mobil

Страница 1: ...AssistX Assist Assist Assist Assist Assist Assist Assist Assist Assist AssistX Mobil INSTRUCTIONS FOR USE US CA AU ...

Страница 2: ... 20389 Patent DE 102015013572 CSS Microsystems GmbH In den Hofäckern 16 74189 Weinsberg Germany www csslabs de 2022 CSS MicroSystems GmbH Please always use the latest version of the Instructions for Use www csslabs de downloads Instructions for use also available in other languages PRELIMINARY VERSION ...

Страница 3: ... Display 10 8 Use of the device 12 Call triggering 12 Confirmation and acknowledge 12 Call Queue 12 Call repeat 13 Out of Range Monitoring 13 Charging 13 Battery condition 14 9 Settings 15 Menu 15 Info menu 15 Languages 15 Subscribers teach in new transmitters to the receiver 15 Set the Alarm signals of the receiver 16 Call repeat 16 Date Time 17 Saving power Power save mode 17 Infrared receiver o...

Страница 4: ...he underlying disease or disability There are basically two types of application AssistX Mobil can be supplemented with a wide variety of control ele ments for individual adjustment to the respective structural or functional disorders and restrictions of the activity These include for example a wide variety of adaptation aids such as buttons proximity sensors sip and puff sensors acoustic sensors ...

Страница 5: ...r make it difficult for caregivers to perceive device alarms through directed movements but also passively through the remaining physical activities For example a strong cough can be detected via the AssistX AS one acoustic sensor With the individually selectable ad aptation aid Sensor AssistX Mobil forms a system for triggering care calls Various sensors and accessories for call triggering are of...

Страница 6: ...6 Instructions for Use AssistX Mobil 3 Explanation of symbols Safety instructions Important instructions ...

Страница 7: ...self The battery can be da maged which can lead to overheating and can lead to injuries The integrated rechargeable lithium polymer battery may only be replaced by authorized personal Mounting The device must be mounted in a way that it cannot fall down Only recommended mounting components from the manufacturer are to be used The user bears the sole risk of mounting The manufacturer is not liable ...

Страница 8: ...ion to people with implanted pacemakers If necessary please contact the device manufacturer of the other active medical device 5 Cleaning and disinfection Before cleaning the device switch it off and disconnect all cables The device should be cleaned daily with a disinfectant and cleanser intended for use in hospitals Cleaning should only be done with a wipe Wipe all surfaces and let them dry or f...

Страница 9: ...X Mobil 9 EN 6 Scope of delivery AssistX Mobil TX transmitter AssistX Mobil RX receiver AssistX Mobil S Charger for transmitter AssistX Mobil Dock for receiver Power supply for Dock USB cable Instructions for Use QuickStart Guide ...

Страница 10: ...0 Confirm OK Button 5 Reset restart 11 Menu up down Button 6 Call Button 12 AssistX Mobil RX Display Transmitter Receiver 7 Device Buttons Ports Display 1 Connected transmitters 2 Charge connected 3 Battery status 4 Out of range warning 5 Call Queue 6 Alarm type 7 Caller Name Time Date Min distance 0 5 m Min distance 1 5 ft Transmitter Receiver 7 2 2 10 6 3 9 4 4 5 5 11 8 12 1 1 1 2 3 5 6 7 4 ...

Страница 11: ...sec until you hear a beep and the LED stops flashing Unplug sensor switch For safety reasons a 6 call is triggered whenever the sensor is plugged off To avoid triggering a call switch off the transmitter before disconnecting the sensor 9 Output socket Alarm signal adjustable via Relay Mode With the Mono Jack Plug you can tap 1 2 with a Stereo Jack Plug 1 2 3 1 Common contact 2 Normally open contac...

Страница 12: ...regiver has to acknowledge the call for the first time now The transmit ter s 7 Confirmation Light and 1 LED then changes from red to green indi cating that the call has been answered by the caregiver In order to end the call in the Call Queue on the receiver the caregiver must confirm the execution of the call to the caller and via the 6 Call Button on the transmitter within an adjustable time Ca...

Страница 13: ...E 0834 a call cannot be deleted The call must always be confirmed first at the receiver then at the transmitter If a transmitter or receiver fails while a call is in progress the other side waits until there is radio contact again and the call sequence can conti nue A call cannot be lost A current call can only be canceled by swit ching the device off and on Note If the range is not sufficient you...

Страница 14: ...on again and needs to be charged Before switching off there will be the following actions Transmitter The transmitter will activate a call alarm and waits until the call is confir med After that it will switch off Receiver Before switching off all active calls have to be served and have to be confirmed socket and connect the micro USB connector to the Dock Then please plug the receiver firmly into...

Страница 15: ...on of the device like firmware version and frequency Languages Call up the menu and choose your language for the menu texts and messanges on the display German English French Subscribers teach in new transmitters to the receiver Up to 32 transmitters can be connected to any number of receivers Adding subscribers Quick Press the Button at the receiver until the LED will light up purple Hold the key...

Страница 16: ...output can activate e g external call signals Aktivate deactivate Alarm signals Go to Menu Settings Alarm Signals Here the signals can be set separately Set the switch mode Trough Menu Settings Relay Mode the switch mode can be set up Relay Mode 1x 1 sec continious Interval Call repeat Here the time can be set when a call should be triggered again if the call has not been finally confirmed at the ...

Страница 17: ...g the display will switch on and will stay on as long as there is an alarm or warning Activate deactivate power save mode Go to Menu Settings Power Save Infrared receiver of the transmitter An infrared receiver is built into the transmitter With a AAC or environment control device a call can be triggered via a built in transmitter This function can be used in addition to a sensor switch The IR rec...

Страница 18: ...an use it to Update the firmware of the transmitter receiver and repeater Assign names eg the caller to the transmitters in the receivers Clone the receiver so that the configuration transmitter names is copied to several receivers Save the receiver configuration Switch the IR channel at the transmitter or switch the IR function on off Setting date and time Instruction by specialist personnel is r...

Страница 19: ...ge to the housing Update Firmware 11 Support If you need any support or for technical questions about the product please contact the reseller where you purchased this product If necessary e g the re seller is not available or cannot help send us an email to support csslabs de Please be sure to include a description of the problem and the serial number of the device Please note If you update a devi...

Страница 20: ...e 0 C to 45 C Transportation and storage 10 C to 50 C 13 Conformity Standards AssistX Mobil meets the requirements of DIN VDE 0834 14 Environmental protection and disposal All electrical and electronic devices must not be disposed with general house hold waste but must be disposed at designated disposal points This product is subject to the European Directive 2002 96 EC and must therefore be pro p...

Страница 21: ......

Страница 22: ... 16387 V2 2 US CA AU PRELIM THE PASSIONATE ASSISTIVE TECHNOLOGY MANUFACTURER ...