COBRA HD | IFUCOBRAHDEN00
INSTRUCTIONS FOR USE
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
29/48
3.2
TECHNICAL DATA
Technical data
Value
Data transfer
USB 3.0
Mains power
External power supply 24 VCC. In: 100-
240Vac - 50/60Hz - 0.9-05A Out: 24Vdc
- 40W
Net cable
with C14 socket
Dimensions (Height x Length x Depth)
300x 400 x 410 mm
Weight
6 kg
Chin rest stroke
70 mm ±1
Minimum height of the chin cup from
the working plan
23 cm
Base movement (x, y, z)
105 x 110 x 30 mm
Working distance
20 mm
Light sources
Technical data
Value
Auxiliary
Led @850nm
Flash
Led @450-650nm
Retinography
Technical data
Value
Spherical ametropia correction
from -20D to +10D (through handle
placed on the joystick)
Image resolution
2448 x 2051 (5 MPixel)
Visual field
50° x 45°
Minimum pupil
2.5 mm
Compatibility with the standard
UNI EN ISO 10940:2009, DICOM (IHE
integration profile EYECARE Workflow)
Fixation points
1 in 1 on the chin rest