Measuring ranges 

 
VA 525 EN V1.01 

 

Page 10 of 41 

 

 

7

 

Flow measuring ranges  

7.1  Flow for different gases 

 

1/4" 

1/2" 

3/4" 

1" 

1 ¼" 

1 ½" 

2" 

Analog 

output 

20mA 

Analog 

output 

20mA

 

Analog 

output 

20mA

 

Analog 

output 

20mA

 

Analog 

output 

20mA

 

Analog 

output 

20mA

 

Analog 

output 

20mA

 

l/min 

[m³/h] 

[m³/h] 

[m³/h] 

[m³/h] 

[m³/h] 

[m³/h] 

Reference DIN1945/ ISO 1217: 20°C, 1000 mbar 

(Reference during calibration)

 

Air 

Low Speed 

25 

20 

45 

75 

140 

195 

320 

Standard 

50 

45 

85 

145 

265 

365 

600 

Max 

105 

90 

175 

290 

530 

730 

1195 

High Speed 

130 

110 

215 

355 

640 

885 

1450 

Adjustment to DIN 1343: 0°C, 1013,25 mbar 

Air 

Low Speed 

25 

20 

40 

70 

130 

180 

295 

Standard 

50 

40 

80 

135 

240 

335 

550 

Max 

100 

80 

160 

270 

485 

670 

1100 

High Speed 

120 

100 

195 

325 

590 

815 

1330 

Argon 
(Ar) 

Low Speed 

45 

35 

75 

120 

220 

305 

505 

Standard 

85 

70 

135 

230 

415 

570 

935 

Max 

170 

140 

275 

460 

830 

1140 

1870 

High Speed 

205 

170 

335 

555 

1005 

1385 

2265 

Carbon dioxide 
(

CO

2

Low Speed 

25 

20 

45 

75 

140 

195 

320 

Standard 

50 

45 

85 

145 

260 

360 

590 

Max 

105 

90 

175 

290 

525 

720 

1185 

High Speed 

130 

105 

210 

350 

635 

875 

1430 

Nitrogen  
(N

2

Low Speed 

25 

20 

40 

70 

130 

180 

295 

Standard 

50 

40 

80 

135 

240 

335 

550 

Max 

100 

80 

160 

270 

485 

670 

1100 

High Speed 

120 

100 

195 

325 

590 

815 

1330 

Oxygen f 
(O

2

Low Speed 

25 

20 

45 

75 

135 

185 

305 

Standard 

50 

40 

80 

140 

250 

345 

570 

Max 

100 

85 

165 

280 

505 

695 

1140 

High Speed 

125 

105 

205 

340 

610 

845 

1380 

Nitrous oxide 
(N

2

O) 

Low Speed 

25 

20 

45 

75 

140 

190 

315 

Standard 

50 

40 

85 

140 

260 

355 

585 

Max 

105 

85 

170 

285 

520 

715 

1170 

High Speed 

125 

105 

210 

345 

630 

865 

1420 

Other gases on request 

 
Please note: 

 
The area outside the pipeline (ambient area of the sensor) must not be an explosive area.

Содержание VA 525

Страница 1: ...VA 525 EN V1 01 Page 1 of 41 EN English Instruction manual VA 525 Compact inline flow sensor with integrated flow straightener...

Страница 2: ...unctioning of the VA 525 are only guaranteed in case of careful observation of the described instructions and notes Sales Office South Gesch ftsstelle S d Zindelsteiner Str 15 D 78052 VS Tannheim Tel...

Страница 3: ...Technical data 7 5 Scaling Analogue output Comprssed Air 8 6 Installation Description 9 6 1 Installation of VA 525 9 7 Flow measuring ranges 10 7 1 Flow for different gases 10 8 Dimensions 11 9 Electr...

Страница 4: ...settings 25 10 3 2 1 Modbus RTU Setup 25 10 3 3 Ethernet Modbus TCP 26 10 3 3 1 1 Network Setup DHCP 26 10 3 3 2 Network Settings static IP 27 10 3 3 3 Modbus TCP Settings 28 10 3 3 4 Modbus Settings...

Страница 5: ...y the material data of the gas component are decisive 2 Safety instructions Please read carefully before starting the device Warning Do not exceed the pressure range of 16 bar Observe the measuring ra...

Страница 6: ...in existing pipelines due to integrated measuring block Sizes 1 1 1 and 2 Integrated flow straighteners no inlet runs necessary Integrated display with presentation of 2 values Flow Consumption Speed...

Страница 7: ...80 C probe tube 20 70 C housing Operating pressure up to 16 bar Power supply 18 to 36 VDC Power consumption max 5W Digital output RS 485 Modbus RTU MBus optinal Ethernet oder Ethernet PoE optional Ana...

Страница 8: ...Speed 0 110 m h VA 525 with integrated measuring block Low Speed 4 20 mA 0 45 m h Standard 0 85 m h Max 0 175 m h High Speed 0 215 m h VA 525 with integrated 1 measuring block Low Speed 4 20 mA 0 75...

Страница 9: ...allation Description 6 1 Installation of VA 525 The sensor VA 525 is pre supplied with the measuring block An installation at customer site is only allowed in the unpressurized state of the system Tig...

Страница 10: ...Low Speed 45 35 75 120 220 305 505 Standard 85 70 135 230 415 570 935 Max 170 140 275 460 830 1140 1870 High Speed 205 170 335 555 1005 1385 2265 Carbon dioxide CO2 Low Speed 25 20 45 75 140 195 320...

Страница 11: ...idth Connection thread L mm B mm H1 mm H mm A mm DN 8 G 1 4 135 55 50 109 1 15 DN 15 G 1 2 135 55 50 109 1 20 DN 20 G 3 4 135 55 50 109 1 20 DN 25 G 1 135 55 50 109 1 25 DN 32 G1 1 4 135 80 80 139 1 2...

Страница 12: ...53 0106 5 m 0553 0107 10 m brown white blue black grey Legend VB Negative supply voltage 0 V Pulse Pulse for consumption VB Positive supply voltage 18 36 VDC smoothed NC Must not be connected to a vol...

Страница 13: ...hable termination therefore the 6 fastening screws from the lid are to be released and set the internal DIP Switch to On It must be ensured that the connection plugs are still plugged and the gasket i...

Страница 14: ...1 page 14 o 41 9 2 4 Ethernet optional PoE Connector plug B Connection cable M12 x coded 8 pole M12 x coded to RJ45 Data LINES 1 2 und 3 4 PoE LINES 5 6 und 7 8 2 8 1 3 5 4 6 7 M12 x 1 Connection cabl...

Страница 15: ...Operation VA 525 EN V1 01 Page 15 of 41 10 Operation Remark Only for version with display The operation of the VA 525 is done by the two capacitive key buttons Up and Enter Up OK...

Страница 16: ...Gas Status in fo SW Version Mo dbus ID Switching to pages 2 5 or back by pressing key Available only with option Pressure AV Time Period for average value calculation could be changed under Sensor Se...

Страница 17: ...otected Selection of a menu item or to change a value is done with the key a final move to the chosen menu item or takeover of the value change needs the confirmation by pressing the key OK Menu items...

Страница 18: ...tings Sensor Setup 10 3 1 1 Input change tube diameter For the VA 525 not adjustable suspended as voted on included measuring section with corresponding measurement block diameter For changes first se...

Страница 19: ...nter via button select the respective position and activate the position with the OK button By pressing the position value is incremented by 1 Complete with OK and activate next number position Confir...

Страница 20: ...erature and reference pressure are 20 C 1000 hPa All volume flow values m h and consumption values indicated in the display are related to 20 C and 1000 hPa according to ISO 1217 intake condition Alte...

Страница 21: ...sing 2x OK Input change of the value by selecting the respective position with button and entering by pressing button OK By pressing the position value is incremented by 1 Complete with OK and activat...

Страница 22: ...Ref Settings Filtertime Setup Sensor Setup Advanced Ref Settings AV Time Under item Filtertime an attenuation can be defined Input values of 0 10000 in ms are possible The time period for averaging c...

Страница 23: ...o point calibration is carried out In case there is already a calibration done it is indicated in menu with calibrated Leave menu with button Back With the low flow cut off activated the flow below th...

Страница 24: ...e If necessary a pressure correction can be made by selecting Set pressure to Select the menu item with the key and then select it with the OK key Enter the corresponding pressure value in the input m...

Страница 25: ...are to be released and set the internal DIP Switch to On Alternatively a 120R resistor can be installed in the plug between pin 2 and pin 4 It must be ensured that the connection plugs are still plugg...

Страница 26: ...fication and description of the Modbus protocol is free to download on www modbus org Supported Modbus commands functions Command Code Description Function Code 3 Read holding register Function code 1...

Страница 27: ...ubnet and Gateway selection keys must be selected and activated with OK The first data field of the selection in this case the IP address is then marked red Confirm with OK the corresponding input men...

Страница 28: ...and then key OK Select the desired position by pressing the and select with OK button Change values by pressing the values takeover by pressing OK Input for the port is done analogue By means of the b...

Страница 29: ...AB Middle Endian 10 3 3 5 Values Register 1001 1500 Modbus Register Register Address No of Byte Data Type Description Def ault Read Write Unit Comment 1101 1100 4 Float Flow in m h R 1109 1108 4 Float...

Страница 30: ...on cf before comma x R 1299 1298 4 UInt32 Consumption Ncf before comma x R 1305 1304 4 UInt32 Consumption kg before comma x R 1311 1310 4 UInt32 Consumption kWh before comma x R 1347 1346 4 Float Velo...

Страница 31: ...put can be defined as pulse or alarm output Selection of field Relay Mode with key and change modus by pressing key OK For alarm output following units could be chosen Nm s m s Nm h m h Nltr h ltr h N...

Страница 32: ...The required password length is 4 digits Please select with button a figure and confirm it with OK Repeat this 4 times With the last figure could be deleted Password input have to be inserted twice C...

Страница 33: ...ightness The actual adjusted backlight brightness is showed in the graph Backlight By activation Dimming after and entering a time a display dimming could be set With Rotate Screen the display informa...

Страница 34: ...und Flow to the channel CH 1 To make changes first select the value item with button and confirm Moving between the different measurements values or to deactivate the 4 20mA with setting to unused by...

Страница 35: ...ed before for value settings Using CLR clears up the complete settings at once For Auto on the max scaling is calculated based on the inner tube diameter max measurement range and the reference condit...

Страница 36: ...Info Settings Info Run Time 2d 21h 23m 12s Vin 23 8V Temp 35 8 92 7 m s 600m h 20 C 53 1 mm 23 C 10 Here you get a brief description of the sensor data incl the calibration data Under Details you are...

Страница 37: ...Value 1 with Unit Consumption m Value 2 with Unit Flow m h Consumption m Value 3 with Unit Gas temperature C All Values could be changed preset in production or with CS Service software Order No 0554...

Страница 38: ...ext recalibration is recommended is internally entered When this date is reached a message appears in the display with the status message CAL Note The measurement will continue without interruption or...

Страница 39: ...rror the sensor has an internal read error on e g EEProm AD converter etc detected I Temp out of Range At media temperatures outside the specified temperature range the status message Temp out of Rang...

Страница 40: ...r the instrument VA 525 On request and additional payment calibration certificates could be issued The precision is given due to use DKD certified flow meters and verifiable 17 Warranty If you have re...

Страница 41: ...components comply with requirements of the following EU directives Elektromagnetische Vertr glichkeit Electromagntic compatibility 2014 30 EU 2014 30 EC RoHS Restriction of certain Hazardous Substance...

Отзывы: