Faults
58
10.2 Possible Faults CHRONOS
®
Fault
Cause
Trouble
shooting
Level too low.
The supply of liquid nitrogen is defective. Open the liquid
nitrogen supply.
Examine the level
and the pressure of
the store tank and
the function of the
valves.
Fault at the level control unit and vessel
control system BIOSAFE-CONTROL
®
ß
See separate op-
erating manual of
BIOSAFE-
CONTROL
®
ß
Level too high, liquid nitro-
gen overflows, the cover
floats
Supply of liquid nitrogen defective
Fault at the level control unit and vessel
control system BIOSAFE-CONTROL
®
ß
Close the supply of
liquid nitrogen,
check the function
of the valves
See separate op-
erating manual of
BIOSAFE-
CONTROL
®
ß
Frost formation on the
vessel
Upper vessel edge thawed This is operational after refilling or cool-
ing-in.
Vessel is directly exposed to strong air-
flow
dry up, if necessary
avoid direct
blowing against the
vessel
Iced-up cover
This is operational after frequent opening If necessary, re-
move the ice care-
fully
Outer vessel iced-up
Loss of vacuum
The vessel is directly exposed to strong
airflow.
Examination /
Reevacuation by
the manufacturer
avoid direct
blow-against of the
vessel
Positive pressure relief
and seal-off device actuat-
ed, vessel extremely iced-
up
Loss of vacuum / pressure inside the
vacuum room
Empty out the ves-
sel / put it out of
operation
Examination / re-
pair at the manu-
facturer’s works
Содержание BIOSAFE 120
Страница 1: ...Operating Manual BIOSAFE 120 220 420 MD Service Hotline 49 0 2741 95 85 75...
Страница 2: ......
Страница 31: ...Safety 27 5 1 Safety Safety advices Handling with cryogenic liquefied gases Source IGV Germany...
Страница 32: ...Safety 28...
Страница 33: ...Safety 29...
Страница 51: ...Accessories 47 660 mm 95 mm 165 mm CHRONOS 220 Towers 77031511 95 x 165 mm...