Level Control Unit and Vessel
Control System
BIOSAFE-CONTROL® ß
17
•
All alarms can be activated and deactivated via a
service menu in the PC software
(1)
. This menu is on-
ly accessible by service personnel. When delivered ex
works all alarms are activated.
Data storage
•
Each device stores data about:
- storage temperature
- operating states (solenoid valve / open cover)
- alarm messages
•
If the system is online, the data is directly transmitted
to the PC
•
Approx. 3 days before the data storage fills up,
a message is displayed. Once the data storage is
full,
the alarm “Memory full“ is triggered and the oldest
data
in the device memory will be overwritten.
Note:
The user can activate and deactivate the alarm
“Memory full” on the device.
When resetting this alarm, an existing alarm
“Memory full” is reset.
If data are not to be saved or no PC is connected,
this alarm should always be deactivated!
When the alarm “Memory full” is deactivated, the
data will still be saved but once the data storage
is
full the oldest data will be overwritten without
previous warning.
The data can be viewed and edited (graphics,
lists, printouts) even when the alarm is deac-
tivated.
Horn off
•
Switches-off the acoustic alarm (no recurrence,
but a new alarm triggers the horn again)
Acknowledge
alarm
•
Acknowledges the alarm / switches-off the
horn / resets the alarm relay
•
If an alarm condition does no longer exist, the
alarm message is erased from the display
(goes to normal operation)
•
If an alarm condition still exists, the alarm
message is also still indicated / horn and alarm
relay are reset for 30 minutes and return after
this period of time
Содержание BIOSAFE 120
Страница 1: ...Operating Manual BIOSAFE 120 220 420 MD Service Hotline 49 0 2741 95 85 75...
Страница 2: ......
Страница 31: ...Safety 27 5 1 Safety Safety advices Handling with cryogenic liquefied gases Source IGV Germany...
Страница 32: ...Safety 28...
Страница 33: ...Safety 29...
Страница 51: ...Accessories 47 660 mm 95 mm 165 mm CHRONOS 220 Towers 77031511 95 x 165 mm...