2.4 Installation / Operation
2.4.1 Installing and Operating the 2083-218#, Frequency Translator Section
1. Connect a -5 dBm to -25 dBm signal to RF IN, J1 (Figure 2.1)
2. Connect the RF OUT, J3, to the external equipment
3. Connect 100-240 ±10% VAC, 47 - 63 Hz to AC connector on the back panel.
4. Set the gain for -10 to +10 dB. Make sure the output stays within -5 to -25 dBm with the gain
selected and the input level provided. (See Section 2.5 Menu Settings).
5. Be sure DS6 (green, DC Power) is on and DS2 (red, Alarm) is off (Figure 2.2).
6. AC Fuse - The fuse is a 5 mm X 20 mm, 2 amp slow blow (Type T) and is inserted in the far slot
in the drawer below the AC input as shown in Figure 2.3. There is a spare fuse in the near slot.
NOTE: If a fuse continues to open, the power supply is most likely defective.
AC Fuse - 2 amp slow blow (Type T),
5 mm X 20 mm
FUSE DRAWER
SPARE FUSE
~
INPUT
100-240± 10%VAC
47-63 Hz
2A MAX
FUSE
TYPE T 2A GDC
250 VOLT
FOR 100 - 240 V~
~
FIGURE 2.3 Fuse Location and Spare Fuse
2083-218# Manual, Rev. 0
Page 10
04/22/14