LPT - Operation Manual
6 INSTALLATION
The sensor in the LPT goes through a burn in period at our factory prior to shipiping so it is ready
for operation upon arrival. If you install the LPT when it arrives, most sensors will not require a
long warm up period (about 5 minutes for Ammonia and Nitrogen dioxide, 2 minutes for the other
gases, except Oxygen which is a minimum of 2 hours). If the device is not installed within two
weeks of delivery, the sensor may require a longer warm up time to stabilize (approximately 48
hours) and provide accurate readings.
NOTE:
CETCI suggests that upon power-up, all sensors be left to warm up for 24 hours
prior to considering the gas readings to be accurate.
NOTE: All sensors are calibrated in the factory and do not require calibration at the time
of a routine installation.
NOTE:
Temperature aff ects calibration. It is important to ensure the gas is at the appropriate
temperature during calibration. If the sensor is being used in an extreme temperature range,
calibration should be done in that same temperature range.
6.1 Special Consider
ations for Diff erent Types of
Sensors
After installing an LPT with an
Oxygen sensor,
leave it to warm up for at least 2 hours before looking
at the readings. If after at least 2 hours, the gas reading is not 20.9%, you should do a respan.
During calibration, when fl owing span gas on an
Ammonia sensor
, if the reading climbs higher than
the calibration point after applying gas for 3 minutes, use that reading as the calibration point. It
should be around 300 ppm.