1.
Connect the power cable to an outlet and press to turn the autoclave on.
2.
Open the door
and
use the measuring cup to pour the correct amount of
distilled water
directly inside the chambe
r
150 ml for Vitale Class 12 and 250 ml for Vitale Class 21.
3.
and close the door correctly.
Load the autoclave with the materials to be sterilized
4.
Press to start,
The autoclave will automatically perform heating, sterilization, depressurization, and drying, according
to the table below data. At the end of the drying process, the autoclave will beep until the operator open the door.
5.
After depressurizing, the autoclave will beep twice every 4 seconds. Slightly open the door for drying.
Note:
if a extra
drying is necessary, see the instruction manual.
6.
At the end of the process, the LED will be on. Wait until the temperature of the material decreases to handle it.
16
USE INSTRUCTIONS - QUICK GUIDE
Read carefully the Instruction Manual
TIEMPO DE
SECADO
*For altitudes above 2.000 m, the sterilization time is 20 minutes. Maximum drying temperature: 129°C
Default Program
Extra Drying
PROGRAMS
STERILIZATION TEMPERATURE
STERILIZATION PRESSURE
STERILIZATION
TIME
DRYING
TIME
HEATING TIME
10 a 60 min.
-------------
1,7 a 1,8 kgf/cm² ou 167 a 186 kPa
---------------------------------
129 a 132 °C ou 264 a 269 °F
--------------------------------
30 min.
20 min.
16 min.*
--------
The Autoclave is an equipment for steam sterilization, it works with high temperature and pressure. The
autoclave must be handled by trained professional and kept at a location with controlled access.
A
ttention! Perform preventive maintenance on the autoclave according to the instruction manual specifications.
Test your autoclave with biological indicators every 7 days.