Incompatible tension level, press to return.
The door was opened, press to return.
Canceled by user.
0/100
0,5/111
1,0/120
1,5/127
1,6/128
1,7/129
LED
CAUSE
Heating time limit, press to return.
Low pressure, press to return.
Over pressure, press to return.
10
e
When the autoclave reaches the sterilization temperature/pressure, it will beep once and the LED
will blink
indicating the sterilization for 16 minutes (table 5).
Obs.:
for altitudes higher than 2000 m, the sterilization will last 20 minutes. If the autoclave does not reach the
proper sterilization temperature in 40 minutes, by lack of water, excess of meterial or air leakage, the cycle will be
automatically canceled. Voltaje oscilations may cancel the cycle at any time.
If a cycle interruption/cancellation is needed, press . In this case, the LEDs
Cancelado
(Canceled) and
0,5
kgf/cm²
will be on, the autoclave will beep twice and return to the standby mode.
At the end of the sterilization, the equipment will beep once, announcing depressurizing. The pressure LEDs will
indicate 0 pressure at the end of this process. After that, the heating/drying LED will blink with 2 beeps every 4
seconds, indicating the need to slightly open the door, for a proper drying (figures 10 and 11).
ATTENTION!
When opening the autoclave for drying, see the correct
position of the door (figures 10 and 11). Keep the handle
all the way down (fig. 11). Vitale Class performs drying
with open door
. It is usual that in this moment, some
steam comes out through the door.
f
ATTENTION!
Even after the cycle is finished, the chamber content will be still at high temperature
Never
directly
touch the chamber, trays or material, wait until the temperature is proper for handling. Remember to use glover for
security and thermal protection.
Wait between 15 and 20 minutes for cooling before start a new cycle
.
After drying, the autoclave will beep once and the LED will keep on. To return to the standby mode, press .
If a extra drying is needed, press to perform this process again.
After use, turn off the equipment pressing for 2 seconds.
g
h
FAILURE INDICATIONS
Table 6
When a cycle is canceled, the Vitale Class autoclave will light up a LED related to the possible cause of
canceling, together with the Cancelado (Canceled) LED, according to the table below: