PHYSICAL INSTALLATION

 

Install the mini-incubator on a flat, leveled and firm surface, at an adequate height for the 

operator to handle it. Leave enough room close to the mini-incubator for the proper handling 
of the biological indicators to be incubated. The place where the mini-incubator will be 
installed must be ventilated, clean and apart from the place where the patients are treated. 
The ideal installation should be done in a separate sterilization and/or material processing 
room.
 

Important! 

Install your mini-incubator where it can be easily unplugged. This is the proper 

way to turn the equipment off completely.

ELECTRICAL INSTALLATION

 

The  voltage  of  the  Cristófoli  Mini-incubator  is  from  90V  to  253V  (automatic  voltage 

selection), as informed in the identification label located under the device (Fig. 3). Make 
sure the wiring voltage where it will be installed is in accordance with the values marked in 
the  field 

Voltage

  of  the  Identification  Label. 

ATTENTION!  Non  compliance  with  this 

requirement  may  damage  your  mini-incubator

Cristófoli  is  not  responsible  for 

damages caused by inappropriate installations and/or voltage

.

  Use a common outlet to plug it in (two pins, Figs. 1 and 1A).

  Never

 use extensions, voltage transformers or any kind of adapters

.

  Your 

Cristófoli Mini-incubator has the following safety devices:

 

The electrical installation must follow the data from the table below:

SAFETY DEVICES

 

LID -

 It closes the incubation area during operation, helping to maintain the temperature 

stable and protect the operator from accidental burns

 (Fig. 2).

 

 -

 

PRINTED FUSE

Built into the electronic circuit, it

s purpose is to protect the building electrical 

wiring against peaks of energy

.

  MAIN UNIT 

- It is made with injected ABS plastic, white on the lower half and transparent on 

the top half (Fig. 2).

  LID  -

  Its  function  is  to  close  the  incubation  area  during  operation  to  maintain  the 

temperature stable (Fig. 2).

  POWER CABLE - 

Used to connect the equipment to the electric outlet (Fig. 3).

  INCUBATION  AREA 

-  Screwed  to  main  unit,  it  is  where  the  indicators  are  inserted  for 

incubation. It has a capacity for 4 biological indicators (Fig. 2).

  LED - 

It indicates when the mini-incubator is on, it also illuminates the incubation area, 

making the visualization of the results easy and clear (Fig. 2).

  IDENTIFICATION  LABEL  - 

Affixed  under  the  device its  function  is  to  identify  the  mini-

incubator technical data (

).

  Fig. 3

MINI-INCUBATOR FEATURES

1

2

HOW TO USE THE CRISTÓFOLI MINI-INCUBATOR

 

ATTENTION! 

Before starting the incubation procedures, the operator must be using latex 

gloves. Handle the biological indicators carefully. The mini-incubator must be installed in a 
proper and exclusive sterilization and/or materials processing room.

 

The symbol 14     appears in the identification label of the mini-incubator, that means the 

selected  item  requires  special  attention  and  that  the  user/operator  must  observe  its 
reference in the instruction manual provided with the equipment regarding the potential 
hazard the device may offer and any actions to be taken should an adverse situation occur. 
Cristófoli Equipamentos de Biossegurança Ltda., is not responsible for any failures and/or 
accidents caused by the non-observance of this warning.

 

ATTENTION! Use only self-contained biological indicators for steam sterilization

 with the 

mini-incubator.  Non-observance  of  this  recommendation  may  cause  incubation  failure, 
poor interpretation of the results and/or damage to the device, as well as loss of warranty.
 

“Correct functioning of sterilization cycles should be verified for each sterilizer by the 

periodic  use  (at  least  weekly)  of  Bis”.  (Centers  for  Disease  Control  and  Prevention. 
Guidelines for Infection Control in Dental Health-Care Settings - 2003. MMWR;52 (No. RR-
17):  page  24).  Each  institution  must  also,  according  to  the  Brazilian  Health  Ministry, 
establish a monitoring routine of its autoclaves, with the objective of detecting faults in 
their equipment and /or process.

 

In  hospitals,  monitoring  must  be  done  on  a  daily  basis  (AORN,  1994).  A  biological 

indicator must be used for every sterilizer load that contains an implantable device. The 
load can only be released after the results of incubation are known.

Use only biological indicators for steam (56ºC)

1 - Plug the mini-incubator into the outlet. The white LED will light indicating the device is 
on;

Note:

 The incubation temperature is 56°C. Wait for the temperature adjustment which will 

take from 40 to 60 minutes;
2  -  After  being  sterilized  according  to  its  manufacturer  instructions,  remove  the  indicator 
from  the  autoclave,  wait  15  minutes,  activate  it  and  insert  it  into  the  incubation  area, 
activate also another indicator from the same batch that was not autoclaved to serve as 
a control;

Note:

 To activate the biological indicator, hold the mini-incubator firmly and insert it half 

way  in  any  of  the  available  incubation  orifices  (Fig.  4),  press  it  carefully  in  any  direction 
until you feel that the internal glass ampoule has broken, bring it back to its due position 
and slide it into the incubation area.
3  -  Wait  for  the  correct  incubation  time  according  to  the  instructions  of  your  biological 
indicator manufacturer (24 or 48 hours);

Note:

 It is important that during the incubation period, the lid is always closed to keep the 

temperature  steady,  open  it  only  when  necessary  (insert  or  remove  the  biological 
indicators).
4 - After the time needed for the incubation has elapsed, remove the biological indicator 
from the incubation area and check the results.

 

It is recommended to examine the biological indicators at regular intervals to verify the 

color change (example, 12, 16, 20, 24 and 48 hours). Bacterial growth is indicated by the 
color change of the biological indicator according to the manufacturer instructions.

Note:

 To cancel the incubation process, simply unplug it from the electric outlet.

 

We suggest that the operator standardize the incubation process and prepare a written 

routine to avoid leaving any requirement behind.

INDICATORS STORAGE

 

The storage of the biological indicators must follow the instructions from the manufacturer, 

specially the temperature exposure limits. The cabinets must be easy to clean, located in a 
dry room with fresh air, free from odors, disinfectants and/or sterilants.

BIOLOGICAL INDICATORS SHELF LIFE

 

The shelf life / expiration date is generally printed on the label and/or box of the biological 

indicators, it is the date after which they cannot be used. We recommend the use of the 

SporTest

” biological indicators which have an 18-month shelf life.

INCUBATION PROCESS MONITORING

 

Monitoring  is  nothing  more  than  controlling  the  incubation  with  self-contained  control 

indicators. The control indicators (indicators that have not been sterilized), have the important 
function  of  evaluating  the  viability  of  the  test  indicators,  as  well  as  the  capacity  of  the 
incubator in providing the ideal conditions to incubate them. In case the result from the 
control is not positive, it is possible that there is some problem with that batch of indicators, 
with their storage, or even with the incubator. Check if a power outage occurred during the 
incubation process. If that is the case, perform the whole test again. That means a new cycle 
must be performed in the autoclave, containing a test indicator and the incubation process 

REQUIREMENTS TO BE OBSERVED FOR THE INCUBATION PROCESS

Fig.1A

Fig.1

Bipolar outlet and plug

(two pins) 10 A.

Fig.2

LID

INCUBATION AREA

BIOLOGICAL INDICATORS (illustrative)

MAIN UNIT

LED (located inside the

incubation area)

POWER CABLE

Fig.4

IDENTIFICATION L ABEL

Fig.3

2

3

02

MODEL

AMPERAGE

VOLTAGE

Cristófoli

Mini-incubator

0,07

90V - 253V AC

automatic

voltage selection

1

  The  storage/installation  must  be  done  in  a  place  protected  from  the  weather  action 
(indoors) in normal conditions of ambient temperature.

  The Cristófoli Mini-incubator can be easily installed. Check if the wiring voltage of your 
building is in accordance with the mini-incubator specifications 

consulting the “

Technical 

Data

” (Page 3) at the time of installation.

 See “

Warranty Certificate

” and “

How to Proceed 

When Service is Needed

” (Page 4).

INSTALLATION INSTRUCTIONS

CRISTÓFOLI

EQUIPAMENTOS DE BIOSSEGURANÇA LTDA

ROD BR 158 Nº 127 - CAMPO MOURÃO - PR - BRASIL

CNPJ 01.177.248/0001-95

INSCR. EST. 90.104.860-65

CEP 87303-327

TEL: 55 44 3518-3434

MADE IN BRAZIL

CAPACITY

4 Indicators

Cristófoli Mini-incubator

MODEL

50/60 Hz

S N

90 - 253V~ - 0,07 kW - 0,1A