User Manual for Handheld Pulse Oximeter
37
9 Troubleshooting
Trouble
Possible Reason
Solution
Unstable
SpO
2
and
Pulse Rate
display
1. The finger is not
placed far enough
inside.
2. The finger is
shaking or the patient
is moving.
1. Place the finger
correctly inside and try
again.
2. Reduce patient
movement.
Unable to
measure
Temperature
1. Temperature probe
is not connected
properly
2. Reinsert the probe into
the device
Device will
not switch
on
1. The batteries are
drained or almost
drained.
2. The batteries are not
inserted properly.
3. The device is
malfunctioning.
1. Recharge or change
batteries.
2. Reinstall batteries.
3. Please contact the local
service center.
No Display
1. The device will
power off
automatically when
there is no signal and
no operation for 1
minute.
2. The batteries are
almost drained.
1. Normal.
2. Recharge or change
batteries.
No Signal
1. Probe off or
incorrect connection
2. Incorrect finger
insert
3. Probe is damaged
1. Reconnect the probe
2. Reinsert the finger
3. Replace a new probe