Database Operational Mode and Area Under the Curve
8-18
Operator's Manual
To capture first alert data, press the HOME
key to display the Main screen, then perform the
following navigation bar options:
Select
YES
or
NO
.
8.6.4
Marking Metrics at the End of a Case
If cerebral oximetry database metrics were not marked during the case, they can be retroactively
captured and calculated.
Note:
Retroactive entry of database metrics can only be entered if within 24 hours of case completion and before
a new case has begun.
Note:
Induction time chosen must be after the time of baselines set.
Retroactively enter the metrics in the following order. AUC will calculate and display automatically
on the Database summary page once the other metrics have been entered.
1.
Baseline (see section
Mark a Baseline at the End of a Case
2.
Induction time (see section
Marking the Induction Time Using Tabular Data
)
3.
Skin closure (see section
Marking the Skin Closure Time in the Database Summary Menu Using Tabular
4.
First alert (see section
Capturing the First Alert Metric
Mark a Baseline at the End of a Case
If no baseline was set during the case, it can be marked at the conclusion. To mark the baseline,
press the HOME
key to display the Main screen. Or, if the monitor has been turned off, turn it
on by selecting the green ON/OFF
key and pressing
PREVIOUS PATIENT
from the Start
screen. Then, perform the following navigation bar options:
BA EVENTS
DATABASE SUMMARY
ALARM AUDIO ON/OFF
NEXT MENU
INDUCTION TIME
SKIN CLOSURE TIME
FIRST ALERT
DOWNLOAD MENU
YES
NO
BA EVENTS
DATABASE SUMMARY
ALARM AUDIO ON/OFF
NEXT MENU
SET BASELINES
SET CHANNEL
EVENT MARK
PREVIOUS MENU
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......