Collecting Metrics in Database Operational Mode
Operator's Manual
8-15
WARNING:
For area under the curve (AUC) to calculate you must set left and right baselines prior to induction
in real-time or accept the auto-baseline value when prompted by the device (approximately 5
minutes after first valid rSO
2
data is received). Baselines should not be reentered at any other time
or the integrity of AUC will be compromised. Similarly, entering skin closure will cease calculation
of AUC.
Note:
For capturing cerebral oximetry metrics, the OR event list should be set (see section
).
8.6.1
Marking the Induction Time
Baselines must be set before the induction time. Induction time is defined as the beginning of the
intubation process, prior to preoxygenation or heavy sedation. To select the current time as
induction time, use the OR event list (see section
Marking the Current Time as Induction Time in the
). To select an earlier time, use the Database summary screen (see section
Induction Time Using Tabular Data
).
Marking the Induction Time Using Tabular Data
To mark the induction event, press the HOME
key to display the Main screen, then perform
the following navigation bar options:
The Enter induction time menu will be displayed, then perform the following navigation bar
options:
Use the
NEXT PAGE/PREVIOUS PAGE
menu options to scroll to the appropriate page and select
SCROLL MENU
. Scroll to the time/rSO
2
value for induction time using the
NEXT/PREVIOUS
menu options. Then, select
SELECT TIME
to enter the appropriate induction time (see
A vertical dashed line and event code will appear on the Main screen at the induction time and
the AUC value in the data channel area will be displayed. The AUC value will be displayed as min-
percent (see
).
BA EVENTS
DATABASE SUMMARY
ALARM AUDIO ON/OFF
NEXT MENU
INDUCTION TIME
SKIN CLOSURE TIME
FIRST ALERT
DOWNLOAD MENU
NEXT PAGE
PREVIOUS PAGE
LAST PAGE
SCROLL MENU
NEXT
PREVIOUS
SELECT TIME
PREVIOUS MENU
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......