− 2 −
Introduction
Product
Model
Features
Multi-Gas Detector
XP-3118
Simultaneously measures combustible gas
concentration in a range of 0% to 10% or 0%
to 100% (LEL) and Oxygen concentration
Multi-Gas Detector (Combustible
gas high-sensitivity type)
XP-3168
Highly sensitive and ideal for the simultaneous
detection of combustible gas concentration in
ppm and oxygen concentration.
Thank you for purchasing the XP-3000 series Multi-Gas Detector. To ensure proper use, please read this
operation manual thoroughly before using the product to prevent gas accidents and oxygen deficiency
accidents, and to conduct maintenance inspections.
Be sure you fully understand this manual regardless of whether you have used a Gas Detector before or
not. Do not use the product in ways not described in the manual.
The following two types of XP-3000 series Multi-Gas Detectors are available.
1. Introduction
Operating conditions:
●
Replace batteries only in area know to be NON-HAZARDOUS.
●
Use the Gas detector in the carrying case specified by New-Cosmos Electric.
●
As a comprehensive measure for the prevention of danger resulting from static electricity.
It is desirable for the user carrying the Gas Detector to wear antistatic clothing and
electrically conductive shoes(antistatic work shoes). The floor should also be electrically
conductive( with an electrical leakage resistance of 10 M
Ω
or less.)
●
In the measurement of the oxygen density, do not use the Gas Detector for any mixtures
other than air mixed with combustible gas or steam mixed with toxic gas.
●
Warning-Substitution of components may impair intrinsic safety
●
Use only Panasonic LR6XJK, Toshiba LR6, or Duracell MN1500 batteries
●
Warning-Do Not Replace Batteries when an explosive gas atmosphere may be present
●
Warning-Potential Electrostatic Charging Hazard-Do not remove leather case in a hazardous
location
●
Warning-Do not connect anything to the USB port when an explosive gas atmosphere may
■
Explanation of Symbols
The following symbols are used for safety purposes:
■
Explosion-proof Precautions
Follow the explosion-proof conditions specified below for operating the Gas Detector.
DANGER
WARNING
CAUTION
: Indicates a hazardous situation that, if not avoided, could result in serious
injury or death.
: Indicates a potentially hazardous situation that, if not avoided, could result
in serious injury or death.
: Indicates a potentially hazardous situation that, if not avoided, could result
in minor injury or physical damage.
: Indicates operational advice and/or instructions.
Note